Dynamic response:
warm-up time
about 30 minutes
90% response time
When the sample airflow velocity is 60 1/h, T90≤ 2 seconds
Calibration:
zero point calibration
Use test gas, process gas without measured components, or alternative gas. Range point calibration
Single point calibration using test gas, process gas with known sample gas concentration, or alternative gas
Single point calibration can be performed using standard gases, and due to the sensor principle used, the zero and range points will not drift separately. According to the corresponding measurement tasks, the zero and range points should be calibrated regularly (recommended once a year)
Materials in contact with the sample gas medium
analyzer
Sensors: gold, silicon nitride oxide;
Sample chamber: stainless steel and acid resistant steel 1.4305;
Sealing: FFKMM75 (perfluororubber) trachea, connector and flame arrester made of stainless steel and acid resistant steel 1.4305, 1.4571