Phone
13810382227
-
Address
303, Building 13, No.1 Chaoqian Road, Changping District Science and Technology Park, Beijing
Beijing Tianyu Wote Instrument Technology Co., Ltd
13810382227
303, Building 13, No.1 Chaoqian Road, Changping District Science and Technology Park, Beijing
1TW-6636 online ammonia analyzerbrief introduction
TW-6636The online ammonia analyzer is a high-precision online automated instrument with functions such as automatic completion of chemical reactions, optical detection, graphic and textual display, control output, and data storage; The instrument adopts a liquid crystal display, which displays measurement results, system information, and a fully Chinese menu operation interface in rich text, charts, and curves; The combination of humanized design concept and high-tech has achieved the performance and product competitiveness of this instrument.
2TW-6636 online ammonia analyzermain features
1) Advanced embedded microcontroller technology;
2) Exquisite structure, disc installation, all aluminum frame box, beautiful and sturdy, strong anti-interference ability;
3) Large screen dot matrix LCD with intuitive and rich display content, capable of switching between single and multiple channels;
4) Adopting the principle of constant pressure dosing, the structure is simple, the measurement accuracy is high, and maintenance free;
5) Adopting 450nm imported light source and photovoltaic cell, with long lifespan, small drift, stability, and reliability;
6) Programmable multi-channel selection design, capable of modifying channel measurement cycles, effectively saving reagents;
7) Equipped with temperature measurement function, it can compensate for measurement data based on temperature;
8) The specific black box function can query historical data, operation records, and calibration records;
3Main technical performance
1) Range: (0-100) µ g/L
2) Display: 320 × 240 dot matrix LCD, Chinese menu
3) Measurement cycle: 5 minutes shorter than usual
4) Instrument indication error: ± 2% F.S
5) Repeatability: not greater than 1%
6) Stability: Baseline drift: using blank calibration, no impact on measurement
Chemical drift: ± 1% F.S/24h (varies depending on reagent stability)
7) Sample conditions: Flow rate: (150~300)mL/min
Temperature: (5-50) ℃
Pressure: (14~140)KPa
The allowed solid content of the water sample is not greater than 5 microns (no gel like substances are allowed)
8) Environmental temperature: (5-45) ℃
9) Environmental humidity: not exceeding 90% RH (without condensation)
10) Reagent type: 1 type
11) Reagent consumption: The longer the measurement cycle, the less reagent consumption;
12) Isolated output: (4-20) mA (isolated output, one per channel)
13) Power supply: AC (85-265) V, frequency (45-65) Hz
14) Power: 60W
15) Dimensions: 690mm x 450mm x 300mm (height x length x depth)
16) Hole size: 645mm × 410mm
17) Weight: 25kg
18) Alarm: Sample breakage alarm, over limit alarm