Overview
Brief description: The TW-5268 infrared spectrophotometer from the manufacturer has a direct reading function for non dispersive measurement results, which can verify the purity of carbon tetrachloride without the need for conversion; Other non carbon chlorine organic solvents such as S-316, tetrachloroethylene, and trichlorotrifluoroethane can be used as extractants;
Product Details

brand TW/Tianyu Water price range Negotiable
Instrument Principle Infrared spectrophotometry Origin category domestic
Instrument type Laboratory type Application field comprehensive

1TW-5268 infrared spectrophotometer manufacturerIntroduction:

This instrument is mainly used for measuring oil content in surface water, groundwater, and industrial wastewater. It can also be used for monitoring oil content in catering industry flue gas and solids. It is a professional instrument for measuring oil content in water bodies, gases, and solids. It has been widely used in fields such as petrochemicals, hydrology and water conservancy, tap water, sewage treatment plants, thermal power plants, steel enterprises, scientific research and teaching in universities, environmental monitoring, and environmental research.

IITW-5268 infrared spectrophotometer manufacturerBasic principle of the instrument:

When a monochromatic light passes through the measured solution, its energy is absorbed, and the intensity of the absorbed light is proportional to the concentration of the measured substance, in accordance with Lambert Beer's law: A=log (1/T)=log (I0/I)=a × b × c.

3、 Main features:

High accuracy: The error of each measurement does not exceed ± 2% (40 ㎎/L oil substance);

Electric modulation light source to improve instrument reliability;

Ø Zero point automatic adjustment, automatic adjustment of instrument full-scale value;

The instrument uses RS232 or USB2.0 interfaces for easy connection with a computer and is compatible with WinXP, Win7, and Win8 operating systems;

Ø It has functions such as mathematical statistics, spectrum display, storage, and printing;

Ø It has the function of measuring instrument calibration coefficients;

Ø It has a direct reading function for non dispersive measurement results and can verify the purity of carbon tetrachloride without the need for conversion;

Other non carbon chlorine organic solvents such as S-316, tetrachloroethylene, and trichlorotrifluoroethane can be used as extractants;

4、 Technical indicators and parameters

baseline drift Less than 0.005AU/60min
Zui low detectable concentration 0.005mg/L (oil concentration in water)
Zui High Measurement Concentration 640000mg/L (oil concentration in water)
Wavenumber range 3400cm-1~2400cm-1(2940nm~4167nm)
wavelength accuracy ±2cm-1
Wave number repeatability ±2cm-1
Range of absorbance 0.0000-3.0000 AU (i.e. transmittance of 100-1% T)
Basic measurement range 0.00~800.00mg/L
repetitiveness Relative standard deviation RSD<1% (for standard oil samples with concentrations ranging from 20 to 80mg/mL)
Linear correlation coefficient R>0.999
Sample analysis time Full spectrum scan takes about 30 seconds
Multiple fixed-point scans About 10 seconds
Non dispersive infrared method About 2 seconds
Zero adjustment of blank solvent Automatic, no need to adjust the fullness
external dimensions 470mm (length) x 330mm (width) x 150mm (height)
host weight 10Kg
Host power supply power 220VAC、50Hz