1TW-2041ND kinematic viscosity testerOverview
TW-2041ND Petroleum Product Kinematic Viscosity Tester, in accordance with GB/T265-1988 "Petroleum Product Kinematic Viscosity Determination Method and Dynamic Viscometer Algorithm", is suitable for determining the kinematic viscosity of liquid petroleum products (Newtonian liquids). This method refers to measuring the time for a certain volume of liquid to flow through a calibrated glass capillary viscometer under gravity at a constant temperature. The product of the capillary constant of the viscometer and the flow time is the kinematic viscosity of the measured liquid at that temperature (the product of the kinematic viscosity at that temperature and the density of the liquid at the same temperature is the kinematic viscosity of the liquid at that temperature), and is applicable to standard GB/T265.
The instrument adopts high-definition color touch screen technology, with rich and clear interface content, fast and flexible touch screen response; Using high-capacity storage chips, it can store up to 2000 measurement results; Using a panel mounted thermal printer, the measurement is completed and the measurement results, sample number, measurement time, viscometer constant, pipe diameter, kinematic viscosity value, and other information are automatically and quickly printed.
2TW-2041ND kinematic viscosity testerInstrument Features
1) 4 viscometers for testing
2) Room temperature~150 degrees, temperature control can be set at any value
3) Can be used as a constant temperature water bath
4) Can be used online with a computer to upload and save information
5) Small size, light weight, easy operation
3technical indicators
1) Display: Equipped with high-definition color touch screen
2) Temperature control range: (room temperature~150) ℃
3) Set temperature: 20 ℃, 40 ℃, 50 ℃, 80 ℃, 100 ℃, any value (above room temperature~150 ℃)
4) Temperature control accuracy: ± 0.1 ℃
5) Test holes: 4 holes
6) Timing range: (0-999) seconds (according to national standards, the flow time of each test sample must be greater than 200 seconds)
7) Timing accuracy: 0.1 seconds
8) Timing times: 1-4 times optional
9) Storage capacity: 2000 test records
10) Printing method: Thermal printer
11) Operating environment temperature: (5-45) ℃
12) Operating environment humidity: (10-80)%
13) Power supply: AC220V ± 5% 50Hz ± 5%
14) Power: 1500VA
15) Dimensions: 45 × 35 × 28 (length × width × height)