- Phone
-
Address
Room 1301, Building B, Changjiang International Building, Mingzhou Road, Beilun District, Ningbo City
Ningbo Qichen Instrument Co., Ltd
Room 1301, Building B, Changjiang International Building, Mingzhou Road, Beilun District, Ningbo City
CE3014N-qvbSurface contour measuring instrumentThe measurement principle is the Cartesian coordinate measurement method, which measures the coordinate points of the surface contour of the measured part through X-axis and Z1 axis sensors. The coordinate point data measured by the sensors is transmitted to the upper PC through electrical components, and the software performs mathematical operations on the collected raw coordinate data to annotate the required engineering measurement items.
CE3014N-qvbSurface contour measuring instrumentMeasurement function:
1. Dimensions: including horizontal distance, vertical distance, linear distance, radius, and diameter.
2. Angle: including horizontal angle, vertical angle, and angle.
3. Position tolerance: including parallelism and perpendicularity.
4. Shape tolerance: including straightness, convexity, and arc.
5. Contour assisted generation: including auxiliary points, auxiliary lines, and auxiliary circles.
Product Atlas:




CE3014N-qvbSurface contour measuring instrumentMeasuring range:
project |
measurement range |
resolution |
sensor |
30mm |
0.02Mm |
drive |
120mm |
0.1Mm |
column |
420mm |
- |
CE3014N-qvbSurface contour measuring instrumentInstrument parameters:
Technical specifications |
Z1Axial accuracy1 |
≤±(0.8+|0.05H|(m)m |
arc2 |
≤±(2+R/8(m)m |
|
angle3 |
≤±2′ |
|
straightness4 |
≤0.5Mm/100mm |
|
Motion Control |
XAxis drive mode |
electric |
ZAxis drive mode |
electric |
|
measure speed |
contourmeasure speed |
0.1-0.5mm/s |
Note:
1. H is the measured height;
2. A standard sphere with a radius of 2mm < R < 10mm, with a sampling arc angle of 120 °;
3. Angle blocks of 60 ° and 90 °, with a sampling length of 5mm at the corner edges;
4. λ c is 2.5mm, measuring speed ≤ 0.3mm/s, measuring flat crystal;
5. The above parameters were measured in an environment with a temperature of 20 ℃± 2 ℃, a step temperature not exceeding 2 ℃/h, and a maximum ground vibration RMS<50Hz 2.5 μ m/s and RMS>50Hz 5.0 μ m/s.
Environmental Requirements
1. There should be no dust, vibration, noise, airflow, corrosive gases, or strong magnetic fields around the indoor area that affect the measurement
2. Vibration Source: Ground Vibration Maximum RMS:<50HZ 2.5 μ m/s/>50HZ 5.0 μ m/s
3. Power supply: AC 220V ± 10% 50Hz, separate ground wire
4. Power requirement: 500W
5. Temperature: 20 ± 2 ℃ Step temperature not exceeding 1 ℃/h
6. Relative humidity: 30% -60% (no condensation)
Instrument configuration
Component Name |
Specification number |
quantity |
ASeries digital contour sensor |
C-Sen-D-A2 |
1 set |
Standard driver |
C-Dri-b1 |
1 set |
Standard column |
C-Col-b1 |
1 set |
Integrated ergonomic machine |
C-Iw-b1 |
1 set |
Electrical system |
1800-D3 |
1 set |
CASContour instrument acquisition and control system |
CAS |
1 set |
Standard adjustment table |
CP-BR-150 |
1 set |
Precision sine flat pliers |
fq-4 |
1 set |
computer |
CP-a1 |
1 set |
printer |
PR-a1 |
1 set |
standard parts |
gauge blockStandard ball |
1 set |
Single section measuring needle measuring rod4*32 |
D-4-32-17 |
1 set |
toolbox |
Tb-10 |
1 set |