Single wire and cable vertical combustion instrument

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

This single wire and cable vertical combustion instrument is suitable for quality inspection/testing units and wire and cable production enterprises for acceptance or qualification review

Product Details

Single wire and cable vertical combustion instrument

this modelSingle wire and cable vertical combustion instrumentsuitUsed for quality inspection/Acceptance or qualification review of testing units and wire and cable production enterprises

The standard recommends two flame calibration methods:

(1) Flame inner flame 50-60mm, outer flame 170-190mm

(2) The time for heating 10g copper head from 100 ℃ to 700 ℃ is 45 ± 5 seconds

If the user uses the second method to verify the flame, a flame calibration device needs to be selected. The flame calibration method is not currently considered for early procurement, and can be added later if necessary.

1Scope of application

This tester is mainly designed for conductors with a diameter greater than0.8mm or cross-sectional areaNoless thanAssess the flammability of a single wire and cable with a diameter of 0.5mm2. Suitable for research, production, and quality inspection departments of electrical and electronic products and their component parts, such as lighting equipment, low-voltage appliances, household appliances, machine tool appliances, motors, power tools, electronic instruments, electrical instruments, electrical connectors, and accessories. This product adopts the recommended product structure for the air mixing structure in Appendix B of GB18380.12-2008, and the details also meet the acceptance requirements of the quality inspection unit.

IIProduct functional features

The vertical combustion test of a single wire and cable is conducted by using a Bunsen burner of a specified size and a specific gas source (propane) to ignite the specimen in a vertical state at a certain flame height and angle. The combustibility and ignition hazard of the specimen are evaluated based on the duration and length of ignition and combustion.

IIIRelated technical parameters

1. Seven inch full-colorProgrammable controller touch screen+PLC control, realizing multifunctional control/detection/calculation/data display

2. High degree of automation: automatic recording of test time, automatic display of test results, automatic timing, automatic ignition, and automatic retraction of the burner after flame applicationYou can choose whether to turn off the gas or not

3. When starting, you can choose whether to automatically ignite;Start, stop, gas, timer, ignition, save, save, and exhaust are all controlled through the touch screen, and the experiment can be completed with a simple touch of your fingertips

4. Timing button coordinationPLC automatically records and stores combustion time consumption

5. The test time system automatically records and generates test results

6. Burner: Inner diameterΦ12mm ± 0.5mm

7. Test inclination angle: 45°The style crossbeam can move forward and backward, and the position of the local lamp can be adjusted according to the thickness of the style;

8. Flame height:Adjustable from 20mm ± 2mm to 190mm ± 1mm

9. Flame application time: 0-999.9s ± 0.1s adjustable

10. Combustion gas95% propane gas (usually replaced by liquefied petroleum gas)

11. Gas: high-purity propane gas or liquefied petroleum gas is used(Note: The gas source and connection head should be provided by the user themselves)

12. Tested wire and cable length:600±25mm;

13. Outer diameter of tested wire and cable: conductor diameter greater than0.8mm or cross-sectional area greater than 0.5mm2

14. This device should be placed without wind and at a temperature of(Test in an environment of 23 ± 10 ℃

15. Standard power of combustion blowtorch:1KW;

16. Power supply voltage:220V/50Hz

17. Gas flow rate:0-1L/min

18. Air flow rate:1.5-15L/min

19. Box material: Steel plate electrostatic spraying (Can)fixedStainless steel material chassis

20. Equipment dimensions: length60 Cm wide 45cm high 156cm

4Applicable standard

GB/T18380.1-2008、GB/T18380.11-2008、GB/T18380.12-2008、

GB/T18380.21-2008、GB/T18380.22-2008

Applicable standard :IEC60332 、 GB/T19666-2005、 GB12972 、 GB13033 、 MT818、MT386、ISO5656