- Phone
-
Address
28 Zhongxin Road, Jiuting Town, Songjiang District, Shanghai
Shanghai Mass Spectrometry Peak Technology Co., Ltd. (Sunny Hengping Instrument)
28 Zhongxin Road, Jiuting Town, Songjiang District, Shanghai
2000 series single beamVisible spectrophotometerFor touch screen type photometer. The new optical and circuit system design ensures that the instrument has high resolution, low stray light, persistent stability, and high signal-to-noise ratio.
Function and Features
4.3-inch touch screen, easy and convenient for users to operate.
Fast wavelength movement speed.
Test the transmittance T and absorbance A of the sample under a single host, automatically set the wavelength, and adjust it to 0Abs/100%.
Coefficient method test: The user inputs the working parameter K, and the concentration result is directly tested and displayed.
● Easy and convenient replacement of lamp source: The flange seat type deuterium lamp structure eliminates the need for special tools to replace the deuterium lamp, eliminating the need for light path debugging steps during lamp replacement, making instrument debugging and maintenance easier.
● Rich optional accessories: Multiple accessories to choose from, providing a wider range of instrument measurement functions.
● Equipped with USB interface.
Introduction to UV Solution Software Features
● Luminescence measurement:Under fixed wavelength conditions, the transmittance T and absorbance A of the sample can be measured, and the data can be exported as an Excel file for users to edit.
Spectral scanning (wavelength scanning):Within the effective wavelength range, set different scanning intervals to scan the absorbance A, transmittance, or energy of the sample, and perform data processing on the spectrum, such as spectrum scaling, smoothing, arithmetic operations, coordinate adjustment, peak/valley automatic detection, and multiple derivative spectra. The records can be printed as standard reports, the data can be exported as Excel files for users to edit, and the spectrum can be exported as BMP image files.
Time scan (dynamic testing): Time scan the absorbance A or transmittance T of the sample at a specific wavelength, setting different sampling time intervals. It can print records as standard reports, export data as Excel files for users to edit, and export graphs as BMP image files.
Quantitative measurement:Standard curves such as straight lines, quadratic curves, third-order curves, and polylines can be tested, and curve parameters can be set. Measurements can be taken at single wavelength, dual wavelength, or triple wavelength, and data can be exported as Excel files for users to edit.
DNA/protein analysis:Quickly measure the absorbance of DNA samples at specific wavelengths, and automatically calculate the concentration and DNA Ratio of DNA/protein. There are two built-in standard testing methods, which can also be extended to other wavelengths and factors for measurement and calculation. The records can be printed as standard reports, and the data can be exported as Excel files for users to edit.
● Multi wavelength measurementCan test sample transmittance T and absorbance A under multiple wavelength conditions, print records as standard reports, and export data as Excel files for users to edit.
3D Map:Can measure three-dimensional spectra.
Audit Tracking:Can provide audit tracking function that complies with GMP regulations (optional)
2000 series single beamVisible spectrophotometerTechnical parameters
Technical parameters | |
model |
UV2000 |
monochromator |
C-T monochromator, 1200 line holographic grating |
detector |
Imported silicon photodiode |
light source |
Imported tungsten halogen lamp 20W/12V, deuterium lamp, light source switching wavelength can be set |
Monitor |
4.3-inch (480 × 272) color touchscreen display |
Spectral bandwidth |
2nm |
wavelength range |
200-1000nm (with a step interval of 0.1nm), can be extended to 190-1100nm |
data interface |
USB |
UV Solution software |
Standard configuration (without GMP audit tracking function) |
Power supply |
220V ± 10% 50Hz 90W / 110V ± 10% 60Hz 90W |
Net weight of the instrument |
10.5Kg |
Gross weight of the instrument |
14.5Kg |
Instrument size |
470×380 × 225 mm |
Package Dimensions |
650×539 × 370 mm |