Welcome Customer !

Membership

Help

Jiangsu Qixing Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Jiangsu Qixing Instrument Co., Ltd

  • E-mail

    m17351886899_1@163.com

  • Phone

    17351886899

  • Address

    No. 286 Tongtai Avenue, Jinhu County Economic Development Zone, Huai'an City, Jiangsu Province

Contact Now

Simple plug-in liquid turbine flowmeter

NegotiableUpdate on 09/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Product Description: QX-LWGY-B Simple Insertion Liquid Turbine Flow Meter is a precision flow measurement instrument, which is divided into tangential sensor and axial sensor according to different sensor forms

Product Details

Product Description:

QX-LWGY-B typeSimple plug-in liquid turbine flowmeterIt is a precision flow measurement instrument, which can be divided into tangential sensors and axial sensors depending on the type of sensor. The plug-in turbine flowmeter sensor (referred to as the sensor) is matched with the display instrument to form an plug-in turbine flowmeter, which can be widely used for measuring the flow rate and total amount of large diameter pipeline source water, circulating water, purified water and other liquids.

1、 Features of plug-in liquid turbine flowmeter sensor:

1. Characteristics of tangential sensors:
Strong anti impurity ability, the tangential impeller can release suspended debris in the fluid at any time during rotation, so that it does not wrap around the blades of the tangential impeller.
Strong resistance to electromagnetic interference and vibration.
The structure and principle of sensors and display instruments are very simple and intuitive, making it particularly easy for users to master their usage and maintenance techniques.
After replacing the impeller and bearings, the instrument coefficient remains unchanged.
Wide flow range, low lower limit flow rate.
The complete flow meter has a small measurement error for the total amount of fluid.
Almost no pressure loss, saving power and electricity consumption.
The sensor can be installed outdoors, and the entire sensor can be submerged in water for a long time.
Sensors with shut-off valves do not require shut-off during installation and disassembly.
Horizontal, vertical, and inclined pipelines can all be used.
The purchase, installation, and decoration costs of complete flow meters are low. 、


2. Characteristics of axial sensors:
The structure and principle of sensors and display instruments are very simple and intuitive, making it particularly easy for users to master their usage and maintenance techniques.
Strong resistance to electromagnetic interference and vibration.
After replacing the impeller and bearings, the instrument coefficient remains unchanged.
Wide flow range, low lower limit flow rate.
The complete flow meter has a small measurement error for the total amount of fluid.
Almost no pressure loss, saving power and electricity consumption.
The sensor can be installed outdoors, and the entire sensor can be submerged in water for a long time.
Sensors with shut-off valves do not require shut-off during installation and disassembly.
Horizontal, vertical, and inclined pipelines can all be used.
The purchase, installation, and maintenance costs of complete flow meters are low.

2、 Insertion type liquid turbine flowmeter sensor flow range:

1. QX-LWGY-B1 tangential sensor flow range:

The measured inner diameter of the tested pipeline is mm

Insertion rod length mm

The flow range corresponding to the nominal diameter DN is m?/h

Full flow range

Accuracy within a flow range of ± 2.5% of the displayed value

Accuracy within a flow range of ± 5% of the displayed value

150

906

13-200

20-200

<20-13

200

23-300

40-300

<40-23

250

36-450

62-450

<62-36

300

52-650

90-650

<90-52

350

70-900

120-900

<120-70

400

92-1100

160-1100

<160-92

500

1106

150-1800

250-1800

<250-150

600

220-2500

360-2500

<360-220

700

280-3500

450-3500

<450-280

800

380-4500

640-4500

<640-380

900

1306

460-5800

800-5800

<800-460

1000

600-7000

990-7000

<990-600

1100

700-8500

1200-8500

<1200-700

2. QX-LWGY-B2 axial sensor flow range:

The measured inner diameter of the tested pipeline is mm

Insertion rod length mm

The flow range corresponding to the nominal diameter DN is m?/h

Full flow range

Accuracy within a flow range of ± 2.5% of the displayed value

Accuracy within a flow range of ± 5% of the displayed value

150

900

7-200

10-200

<10-7

200

12-300

20-300

<20-12

250

18-450

31-450

<31-18

300

26-650

45-650

<45-26

350

35-900

60-900

<60-35

400

46-1100

80-1100

<80-46

500

1100

75-1800

125-1800

<125-75

600

110-2500

180-2500

<180-110

700

140-3500

225-3500

<225-140

800

190-4500

320-4500

<320-190

900

1300

230-5800

400-5800

<400-230

1000

300-7000

495-7000

<495-300

1100

350-8500

600-8500

<600-350

3、 Main technical parameters of the plug-in liquid turbine flowmeter sensor:

Product Structure Diagram
Instrument model QX-LWGY-B1 QX-LWGY-B2
Disassembly and assembly method Disassemble and assemble after cutting off the flow No need to cut off the flow, the ball valve can be disassembled and assembled after closing it
accuracy ±2.5% ±2.5%
monitor optional optional
output signal Optional: Pulse or 4-20mA Optional: Pulse or 4-20mA
communication interface Optional RS485 Optional RS485
power supply +24VDC±15% +24VDC±15%
Explosion-proof rating Optional ExdIIBT6 or ExiaICIT4 Optional ExdIIBT6 or ExiaICIT4
Protection level IP65 IP65
Total power consumption <1W <1W
Instrument material 304SS 304SS
Nominal Pressure 1.0MPa 1.0MPa
measurement range -20℃~120℃ -20℃~120℃