Shanghai Yidian IoT SGW X-5 Microscopic Melting Point Analyzer

NegotiableUpdate on 03/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Shanghai Yidian Wuguang SGW # 174; X-5 Micro Melting Point Analyzer Company currently has over 100 professional physical optical instrument products in 8 series, including polarimeter, melting point analyzer, refractometer, haze analyzer, colorimeter, laser particle size analyzer, turbidity analyzer, stress analyzer, etc.
Product Details

Shanghai Yidian Wuguang SGW ® X-5 Microscopic Melting Point Analyzer

Adopting PID temperature regulation technology to accurately control the changes in measured temperature.

By using a microscope camera and a color LCD touch screen, the melting process of the sample under the microscope is visually displayed on the screen, allowing users to more accurately observe and control the measurement process.

The measured temperature values (initial melting, final melting) can be stored for users to easily view at any time.

The test results can be transmitted to a PC for processing through a USB interface, or connected to a micro printer to print the test results.

The instrument has two melting point measurement methods, capillary method and hot table method, which can measure solid and powder samples to meet different user needs.

Shanghai Yidian Wuguang SGW ® X-5 Microscopic Melting Point Analyzer

Instrument model

SGW ò X-5

measuring range

Room temperature~360 ℃

measuring method

see

Minimum indication

0.1℃

heating rate

0.2℃/min、 0.5 ℃/min, 1.0 ℃/min, 1.5 ℃/min, 2.0 ℃/min

3.0℃/min、 4.0 ℃/min, 5.0 ℃/min

Indication error

Capillary method

≤200℃:±0.4℃; >200℃:±0.7℃

Hot table method

Room temperature~100 ℃: ± 0.4 ℃; 100℃~200℃:±0.6℃; 200℃~300℃:±0.8℃

Repeatability of indication values

When the heating rate is 1.0 ℃/min, the standard deviation is 0.3 ℃

Linear heating rate error

±10%

Capillary size

outside diameter?1.4mminside diameter?1.0mmLength 90mm

Sample height

3mm

Standard slide ruler

Glass slide: 76.2mm x 25.4mm x 0.8mm; Cover glass: 18.0mm x 18.0mm x 0.17mm

Field of view diameter

?2.0mm

Data storage

1000group

Display mode

7Inch LCD touch screen

Calibration method

Multi point calibration

communication interface

USB(Connected to the upper PC), USB flash drive

Printer Model

CK NT-T24S(Optional)

power supply

220V±22V, 50Hz±1Hz

Instrument dimensions

310mm×230mm×505mm

Net weight of the instrument

4Kg