- Phone
-
Address
No. 58 Sanyang Road, Baoshan District, Shanghai
Shanghai Mitong Electromechanical Technology Co., Ltd
No. 58 Sanyang Road, Baoshan District, Shanghai
Shanghai Mitong is a high-quality supplier of flash point instruments for fully automatic pesticide productsIt is designed and manufactured according to the requirements specified in the national standard NY T 1860.11-2016 "Guidelines for Determination of Physical and Chemical Properties of Pesticides Part 11 Flash Point". Suitable for determining the flash point of pure liquid pesticides, active ingredients, and formulations in pesticide registration tests using the methods specified in this standard.
1、 Main functional features
1. SYP-1860D typeShanghai Mitong is a high-quality supplier of flash point instruments for fully automatic pesticide productsUsed to determine the closed flash point value of pesticide products.
2. Adopting an 8-inch pure industrial true color touch screen, the interface can switch between Chinese and English freely, and the Windows CE embedded operating system.
3. Equipped with instrument self diagnostic function to check for faults and display their status.
4. Automatically calibrate the influence of atmospheric pressure on the experiment and calculate the correction value.
5. The lower computer stores 120 sets of historical data
6. The ignition mode can be manually or automatically selected at will, and this machine uses electric ignition.
7. Detection, lid opening, ignition, printing, fully automated.
8. Using Renesas microcontroller. Imported PT1000 high-precision temperature sensor from Germany.
9. It has the function of automatically stopping work to prevent fire and overheating.
10. Built in high-precision flash point detection sensor, with detection time up to milliseconds.
11. Simulate and track the function curve of temperature rise and test time.
12. This instrument has a software copyright certificate (certificate number: 6088745).
2、 Main technical parameters
1. Working power supply: AC (220 ± 10%) V, 50Hz;
2. Temperature measurement: Range: 0-300 ℃;
Temperature display unit: 0.1 ℃;
3. Repeatability: Meets the requirements of NY T 1860.11-2016 standard.
4. Reproducibility: Meets the requirements of NY T 1860.11-2016 standard.
Temperature display unit: 0.1 ℃;
5. Sample heating rate: Step A: (5-6) ℃/min;
Step B: (1-1.5) ℃/min;
6. Mixing speed: ⑴ (90-120) revolutions per minute, suitable for test step A.
⑵ (250 ± 10) revolutions per minute, applicable to Test Step B.
7. Ignition method: Ignite using electronic ignition method;
8. Experimental conditions: Environmental temperature: (10-40) ℃; Relative humidity: ≤ 80%;
9. Power consumption of the whole machine: not exceeding 500W.
10. Dimensions: 520mm x 360mm x 310mm (length x width x height) when the lifting arm is not lifted;
(When the lifting arm is lifted) 520mm × 360mm × 420mm (length × width × height)