- Phone
-
Address
1st Floor, No. 10, Lane 255, Xiaotang Road, Fengxian District, Shanghai
Shanghai Xinsong Industrial Co., Ltd
1st Floor, No. 10, Lane 255, Xiaotang Road, Fengxian District, Shanghai
WTW Online SensoLyt ® 700IQ ORP composite electrode sensor PtA-109125
SensoLyt 700IQ pHsensorWTWItem No.109170
SensoLyt ® Introduction to PtA-109125 Sensor for 700IQ PH Meter
SensoLyt 700IQ pHSensors are composed ofWTWProduced by the company, it has self-monitoring function, can indicate glass electrode damage, and provide a sturdy mechanical shell to protect the electrode. This sensor has a preamplifier function andNTCTemperature sensor ensures stable operation in harsh environments. The sensor components include different types of composite electrodes, suitable for monitoring various water bodies, from heavily polluted wastewater to drinking water. In addition, the sensor supports multiple calibration methods and has a built-in lightning protection circuit, which complies with multiple international standards.
SensoLyt 700IQ pHsensorWTW 109170
1、 Characteristics
1Sensor self-monitoring function, indicating glass damage;
2A sturdy mechanical casing effectively protects the glass electrode;
3Replacing the glass electrode is very easy;
4Removable correction.
Test scope:±2000 mV
WTW Online SensoLyt ® 700IQ ORP composite electrode sensor PtA-109125Scope of application:
1Municipal and industrial wastewater
2Suspension, emulsion
3Containing protein andS2-The medium
(3)SensoLyt SEA-HP
Resistant to higher pressurepHComposite electrode
Test scope:pH 4-12
SensoLyt ® 700IQ ORP Composite Sensor PtA-109125 Scope of Application: Pipeline Installation
(4)SensoLyt DWA
Long service life, more accurate testing, suitable for low conductivity applications such as drinking water.
SensoLyt ® 700IQ PH meter sensor PtA-109125 testing range:pH 0-14
Scope of application: Drinking water
(5)SensoLyt ECA
economypHComposite electrode, with1A pinhole shaped membrane, gel electrolyte. More commonly used in sewage monitoring.
Test scope:pH 2-12
Scope of application: Ordinary sewage
model:SensoLyt® 700 IQ(SW*)
Measuring principle:Use a composite electrode for potential measurement; Built in microprocessor, two core shielded cable containing power and data transmission
Measurement range and resolution
unit of measurement,measurement range,resolution
pH 0.01 pH
0.00 . .. 14.00 pH (Depends on the electrode)
ORP-2000 mV . .. +2000 mV
1 mV
temperature measurement
NTC
measurement range - 5°C . .. + 60°C
precision±0.5 K
resolution0.1 K
reaction timet99 < 15 s
temperature compensation:0 °C . .. + 60°C
Electrode inspection detectionPHIs the electrode glass damaged
Store calibration data have
WTW online SensoLyt700IQ ORP composite electrode sensor PtA-109125Applicable electrodes SEATheGDATheECAThePtA
surrounding environment Operating Temperature0 …60℃, storage temperature:-5 …65℃
material Electrode tip:PVC,Electrode rod:316Tistainless steel,Protection level:IP 68
pressure resistance maximum10Ba (including sensor connection cable)
insertion depth minimum40cm, the largest100m
Flow rate requirements No flow rate requirements
Lightning protection Built in lightning protection circuit
EMI/RFI:conform toEN61326 Class B, FCC Class AStandard stipulates
certification:CE, cETLus
power consumption:0.2tile
size:508×40 mm, L×D
515×59.5 mm, LXD(SW)
weight(cable-free) about970g, SWabout1800g