The semi micro Ubbelohde viscometer $r $n automatic Ubbelohde viscometer consists of HCT series high-precision constant temperature bath, IV viscosity measurement unit, CL automatic cleaning module, etc;
Semi micro Ubbelohde viscometer
1、 Overview
1. Manual liquid inlet, automatic measurement, automatic cleaning, drying, and soaking functions;
2. Say goodbye to the era where viscosity tubes were consumables and to manual cleaning.
Semi micro Ubbelohde viscometer
2、 Product Features
1. Adopting modular design: The automatic Ubbelohde viscometer consists of HCT series high-precision constant temperature bath, IV viscosity measurement unit, CL automatic cleaning module, etc;
2. The viscosity measurement unit and automatic cleaning module adopt a one-to-one approach to avoid mutual interference during the measurement process;
3. Adopting intelligent near-infrared photoelectric sensors and special detection methods, it can meet the measurement of samples of various colors (including pure black masterbatch);
4. Measurement methods: single point method, serial extrapolation method, parallel extrapolation method;
5. The constant temperature bath adopts a dual chamber principle design and circulation mode, ensuring that the temperature fluctuation and uniformity reach ≤ 0.01 ℃;
6. Safety: The instrument has high and low bath level alarm functions and over temperature protection function;
7. The temperature control part of the bath and the body of the constant temperature bath can be easily separated, making it convenient to clean the mixing area of the bath;
8. The communication between each module is connected using standard UBS data cables, which is very convenient to connect;
9. Adopting stainless steel armored fiber optic system, which has the advantages of corrosion resistance and long service life;
10. Adopting a high-pressure pulse reciprocating automatic cleaning method, even the most difficult polymers can be easily cleaned.
11.Drying method, while ensuring the drying effect and efficiency, can use more environmentally friendly and safer drying reagents for drying;
12.The classification and collection function of waste liquid can avoid adverse chemical reactions and facilitate the treatment, recycling, and reuse of waste liquid;
13. Viscosity measurement software can display results: intrinsic viscosity value (ultimate viscosity), relative viscosity value, specific viscosity value (viscosity), Fickentscher's K value, viscosity average molecular weight value, kinematic viscosity value, polymerization degree, dynamic viscosity value, measurement standard/absolute/percentage deviation value, specific viscosity value (specific viscosity), time, and their average values;
14. The V1.0 version of the Zhuoxiang automatic viscosity measurement software provides the vast majority of viscosity calculation formulas currently used internationally: Schulz Braschke, Hagens Huggins, Solomon Xiuta, Billmeyer, Maron, Martin, average molecular weight Mark Houwink, kinematic viscosity, dynamic viscosity, Fickentscher's K value, relative viscosity, specific viscosity (viscosity), specific viscosity (viscosity), and specific viscosity (viscosity)Free addition of self provided calculation formulas for customers;
15. The software has permission level management function to ensure that important experimental data is not arbitrarily modified. Data can be queried, previewed, exported to Excel, printed, and statistically analyzed.
3、 Main technical parameters of the instrument
Measurement time range |
0.01-999.99 (s) (expandable to 9999.99S) |
Measurement timing accuracy |
≤0.01s(Actual resolution: 0.0001s) |
Viscosity measurement range |
0.3-5000 (mm2/s) (expandable to 50000 mm2/s) |
Test temperature range |
15~65℃ |
Temperature field accuracy and uniformity |
≤±0.01℃ |
Usage of cleaning and drying solution |
<13ml (single dose) |
Measurement digits |
4 people |
Number of cleaning positions |
4 people |
Applicable viscosity tubes |
Ubbelohde viscosity tube, miniature Ubbelohde viscosity tube, diluted Ubbelohde viscosity tube
(Compliant with ISO3105, GB/T1632-93, ASTM D 446, DIN 51562 standards)
|
Applicable test samples |
PBAT、PET、PETG、PBT、PBS、PA、HTPA、PPA、PES、PVC、PAM、PAA、PVDF、ACR、POM、PAAS、HA、PLA 、PGA、PLGA、PCL、PGCL、PMMA、PC、MQ、FFKM、PEEK、PEKK、ECDP、PS、PAN、PEN、PHA、PVP、PTT、 Polymer materials such as chitin, cellulose, polysaccharides, and organosilicon |
Applicable standard |
ISO1628 (excluding 1628-3) ASTM D 4603、ISO307、ASTM D 789、ISO1157、GB/T14190、GB/T17931、GB/T3401、GB/T12006、GB17514、GB/T12005、GB/T28610、GB/T1632、 HG/T2234、HG/T3604、HG/T2838、HG/T 2758, viscosity testing standards in pharmacopoeia, etc. |