The semi-automatic amino acid analyzer (economical type) has improved the disadvantages of high procurement, use, and maintenance costs of imported instruments, and better meets the application needs of domestic users.
Product Introduction:
The HT-1105 semi-automatic amino acid analyzer (economic type) independently developed by Qingdao Haitai Yinuo Technology Co., Ltd. draws on mature international technology and design concepts, optimizes and improves the cost of imported instruments in procurement, use, and maintenance processes, and better meets the application needs of domestic users. Multiple performance indicators of the instrument have reached the industry's mature level.
The HT-1105 semi-automatic amino acid analyzer (economical) can analyze more than 20 hydrolyzed amino acids and 43 free amino acids, and can adjust the program according to user needs to achieve functions such as minority amino acid analysis or rapid analysis.
Detailed technical parameters:
1. Retention time reproducibility: All RSD ≤ 0.1%, Arg 0.1%;
2. Peak area reproducibility: All RSD ≤ 0.5%, Gly<0.5%, His<0.5%;
3. Detection limit: All<3pmol (S/N=2), Asp ≤ 2.5pmol, His ≤ 1pmol;
4. Separation degree: All hydrolyzed amino acids ≥ 98% (average), Thr Ser 100% for threonine/serine, and Gly Ala 90% for glycine/alanine;
5. The separation column adopts a high theoretical plate number separation column to improve analysis efficiency and save reagent consumption;
6. Built in full channel vacuum degassing machine, with good degassing performance;
7. Adopting an efficient post column derivatization reactor, the reaction is good and the time is short;
8. Pre aerate the reaction solution and buffer solution with nitrogen to remove oxygen, and protect them with positive pressure nitrogen gas;
9. Buffer solutions and reagents are stored in independent reagent bottles at room temperature, with a shelf life of 1-2 years;
10. High precision dual plunger reciprocating pump: strong corrosion resistance and excellent stability;
11. Column temperature box method: semiconductor refrigeration and heating, which can achieve temperature gradient separation function;
12. Post column derived unit: electric heating, high efficiency, safe and reliable;
13. Detection unit: simultaneous detection of dual wavelengths at 570nm and 440nm;
14. Management System: Compliant with GLP/GMP regulations, capable of controlling instruments, collecting and processing data, and implementing automatic cleaning programs for the system; Capable of conducting system verification tests, customizing customer reports, and ensuring user safety; It has functions such as automatic recording, fault location, and remote network assistance;
15. Can provide software in Chinese or English versions;
16. Buffer reagent kit: independently developed and formulated domestically to reduce usage costs;
17. Service content: sample testing, method establishment, specialized training, regular follow-up visits;
*The product configuration can be adjusted according to customer needs, and specific parameters are subject to the actual product!