DANI, an Italian company, has played an important role in the field of gas chromatography since 1975 and has always been at the forefront of gas chromatography pretreatment, including static headspace samplers, dynamic headspace samplers, fully automatic thermal analyzers, etc.
In the past six years, DANI has achieved a market share of 2/3 in the headspace sampler market, becoming the leading manufacturer in the global headspace sampler market. Taking into account the opinions of all experts and *'s technology, DANI's * new product HSS 86.50 has been launched. The technology of DANI HSS 86.50 has demonstrated the best advantages of headspace in any application field.
HSS 86.50 utilizes DANI's "valve and loop" technology, which was first proposed and widely praised by users. Under different back pressures in the system, this technique ensures that the amount of gas entering is constant and can be repeated. The relative standard deviation is less than 1%.
The DANI HSS 86.50 headspace sampler is a completely independent system that can be used with all types of GC. The interface connection is fast and easy. Just connect the carrier gas and insert the delivery pipeline into the GC injection port. All parameters can be displayed clearly through the keypad. The original imported DANI Hss86.50 headspace sampler from Italy adopts quantitative ring volume injection; 6 bottle positions are stacked for preheating, effectively improving work efficiency, and 44 bottle positions are continuously analyzed, greatly saving labor costs.
DANI HSS 86.50PlusTechnical specifications of headspace sampler
Sample bottle (VIALS)
Sample bottle quantity: 44
Sample bottle material: Neutral glass
Sample bottle volume: 20ml; 10ml (with seat cushion)
Sample bottle gasket: butyl rubber with polytetrafluoroethylene layer (PTFE); Silicone rubber sample bottle cap with PTFE layer
Heating furnace (OVEN)
Temperature range: 15 ℃ to 200 ℃ above room temperature with an accuracy of 1 ℃
Accuracy: 0.5% of full scale
Stability: ± 0.5 ℃
Heating method: electronic, with ventilated composite area (MANIFOLD)
Temperature range: 15 ℃ to 200 ℃ above room temperature with an accuracy of 1 ℃
Accuracy: 1.0% of full scale
Stability: ± 0.5 ℃
Heating method: electronic
Sampling valve: MV 65106HT, six way valve
Sample quantification ring: 0.5ml volume, 1ml volume of nickel tube, 3ml volume of nickel tube, nickel tube (optional)
Transfer Tube Temperature Range: 15 ℃ to 220 ℃ above room temperature with an accuracy of 1 ℃
Accuracy: 2.0% of full scale
Stability: ± 0.5 ℃
Heating method: electronic
Heating furnace turntable (OVEN CROUSEL) heating hole quantity: 6
Shake: SOFT, slight shake, motor stride: 100 steps/second FAST, vigorous shake, motor stride: 200 steps/second
Programmable microprocessor: 8-bit
Control keys: START, RESET, LOCK, SET, ACT, UP, DOWN
Operation indicator lights: WAIT, READY, AUTOSTART, RUN, LOCK
Valve time buttons and corresponding indicator lights: PROBE, PRESS, VENT, VALVE, TEST display (PLAY) 16 digit word count display
Gas Control (GAS) Carrier Gas: Pressure Control Gauge with Digital Pressure Reader (provided to Part-No.0310.100001 and)
Part.No.0310.100002)
Flow control gauge with digital pressure reader (provided to Part. No. 0310.100003) Auxiliary gas: Pressure control gauge with digital pressure reader (provided to all versions)
Control Interfaces RS232 serial communication BCD sample number output
AUTODIAGNOSTIC TEST RAM ROM
Heating furnace probe, valve probe, transmission tube probe
Common indicators of carrier gas pressure and auxiliary gas pressure (UTILITIES)
Power supply: 220V (± 10%), 50Hz, 600VA, 110V (± 10%), 60Hz, 600VA
Safe tube: 3.15AT, 250V (220V) 6.30AT, 150V (115V)
Dimensions and Weight
Size: 42cm wide x 47.4cm high x 61.7cm thick
Weight: 32Kg
Updated on September 2, 2020 at 10:22:28