SGSY 1180 Multi functional Microplate Reading Machine,Provided solutions for users with diverse needs,can doFluorescence based on filter monochromator system(Top Reading), can be applied toshine(glow/Flash absorption light detection,There are also continuous spectrum absorption light detection based on grating spectral detectors,To provide researchers with a multifunctional microplate reader with excellent sensitivity and performance, and powerful upper computer software to assist users in instrument control, various types of data detection and analysis.The special dual system design ensures sensitivity during user detection,It's also very convenientFlexibility to meet the different wavelength requirements of users for different experiments, equipped with a liquid separator, upgraded on the basis of luminescence detection to meet the needs of flash detection.Versatile and applicable to most daily applications,Meet the requirements of high performance and multifunctionality in various application scenarios.
SGSY 1180 Multi functional Microplate Reading MachineProduct Features:
★ Powerful data detection and analysis capabilities
(1)Easy to use can save time on training, data generation, and data analysis, and obtain results faster.
(2)support6-384Orifice plate detection, up to3Various detection functions, including full wavelength light absorption, fluorescence, and chemiluminescence, supporting constant temperature incubation and oscillation.
(3)A multifunctional instrument that can quickly and easily run the required applications according to user needs, and has powerful data analysis capabilities.
(4)Just need1-2-3Step by step, easily complete detection and analysis, instrument settings-detection and analysis-Obtain results.
(5)Free cloth boardFree detection steps.
(6)Equipped with stable delay function to solve the impact of liquid level shaking on detection results.
★fluorescence
(1)The excitation light source is an imported halogen tungsten lamp, and a xenon flash lamp can be optionally equipped as the excitation light source.
(2)The dynamic range of automatic gain.
(3)Adopting intelligent gain conditions, adjust the amplification factor according to the sample intensity.
(4)Normalize samples of different strengths and batches to the same level for calculation and analysis.
(5)Equipped with a light source reference detector to ensure the stability of the excitation light source.
(6)Configure and optimize bandwidth filters, andPMTThe stability and sensitivity of the detector have been further optimized and improved.
★ Chemistryshine
(1)adoptPMTThe detector can effectively meet the sensitivity requirements of various types of luminescence experiments and meet the requirements for flash experiments
(2)Flash detection: after adding reaction reagentsstandPerform testing immediately to meet the testing needs of flash type luminous systems, and complete the testing before the decay disappears in one or a few seconds.
(3)Special liquid level separation design allows for simultaneous liquid separation and detection in chemiluminescence measurements without the need for movement.
★ absorb light
(1)Featuring a specially optimized dual system design.
(2)An absorption light detection system using a tungsten lamp as the light source (optional xenon flash lamp) and a filter monochromator. This system has12通道PDDetector, among which12The channel is a sample detection channel,1Each channel is a reference detection channel. Can detect simultaneously12The samples from each channel ensure both detection speed and stability.
(3)A continuous spectral absorption light detection system using a xenon lamp as the light source and a grating as the monochromator. The system adoptsCMOSAs a detector, it can quickly perform190~850nmAbsorption spectrum scanning,1nmStep forward, you can1sPerform UV testing on a single pore internally-Visible absorption spectrum scanning.
★Liquid separation function
(1)Specially designed liquid separation function, standard configuration is a single liquid separator, and can be expanded to multiple configurations according to user needs3A liquid separator.
(2)The instrument is equipped with three separation holes, which can simultaneously place three types of reagents for automatic separation.
(3)Liquid separator separation range4~995ul,1ulStep by step, it can meet the liquid separation requirements of different doses for various types of experiments.
★Oscillation temperature control
(1)Oscillation: linear oscillation, continuous oscillation, and interval oscillation. The oscillation speed, intensity, and oscillation time can be adjusted according to demandlineset up.
(2)Temperature control: room temperature+3℃-65℃, can enter the samplelineConstant temperature control, and can be set to start detection only after reaching the specified temperature.
★Optional audit trail function
Hierarchical permission management meets the demand for traceability of electronic records.
★ PartCommon applications
(1)Absorption of light:ELISA/Immunoassay methodDNA/RNAQuantification, enzyme kinetics, bacteria/Microbial growth, protein quantification(BCATheBradfordTheLowrydetermination method)
(2)Fluorescence: Calcium andcAMPdetection(GPCR)Cell viability and cytotoxicity, fluorescent protein quantification, nucleic acid quantification
(3)Luminescence: Cell viability and cytotoxicity, Mycoplasma detectionNanOBRET/BRET, Single/Dual luciferase reporter gene detection
Technical Specifications:
model |
SGSY-1180A |
SGSY-1180B |
fluorescence | ||
detector |
Photomultiplier tube(PMT) |
Photomultiplier tube(PMT) |
light source |
tungsten-halogen lamp |
tungsten-halogen lamp |
320nm-1000nm |
320nm-1000nm |
|
Emission wavelength range |
360nm-800nm |
360nm-800nm |
Excitation filter |
standard configuration3Group(355nmThe485nmThe544nm) |
standard configuration3Group(355nmThe485nmThe544nm) |
Detecting filter |
standard configuration3Group(460nmThe538nmThe590nm) |
standard configuration3Group(460nmThe538nmThe590nm) |
Wavelength bandwidth |
30nm |
30nm |
Wavelength accuracy |
<0.2nm |
<0.2nm |
Fluorescence sensitivity |
Top Reading<2fmolFuorescein/well |
Top Reading<2fmolFuorescein/well |
Detection order of magnitude |
Top Reading>6An order of magnitude |
Top Reading>6An order of magnitude |
Enzyme linked immunosorbent assay (ELISA) plate |
support6-384orifice plate |
support6-384orifice plate |
chemiluminescence | ||
detector |
Photomultiplier tube(PMT) |
Photomultiplier tube(PMT) |
wavelength range |
270nm-670nm |
270nm-670nm |
sensitivity |
<40amoLATP/well |
<40amoLATP/well |
Detection order of magnitude |
> 6An order of magnitude |
> 6An order of magnitude |
Enzyme linked immunosorbent assay (ELISA) plate |
support6-384orifice plate |
support6-384orifice plate |
absorb light | ||
system |
12Channel optical path detection system |
12Channel optical path detection system |
light source |
Tungsten lamp (optional xenon flash lamp) |
Tungsten lamp (optional xenon flash lamp) |
detector |
photodiode |
photodiode |
wavelength range |
190nm-1000nm,12Position filter; standard configuration405nmThe450nmThe492nmThe650nm |
|
Wavelength accuracy |
±1.0nm |
±1.0nm |
Wavelength repeatability |
<0.2nm |
<0.2nm |
linear range |
0.000-4.000Abs |
0.000-4.000Abs |
Enzyme linked immunosorbent assay (ELISA) plate |
only support96orifice plate |
only support96orifice plate |
absorption spectrum | ||
system |
/ |
Full wavelength system |
light source |
/ |
xenon lamp |
detector |
/ |
2048-CMOSdetector |
wavelength range |
/ |
190nm-1000nm |
Wavelength accuracy |
/ |
±1.0nm |
Wavelength repeatability |
/ |
<0.2nm |
linear range |
/ |
0.000-1.500Abs |
Enzyme linked immunosorbent assay (ELISA) plate |
/ |
support6-384orifice plate |
spectrogram |
/ |
Real time output of full band spectra for each hole |
autosampler | ||
quantity |
standard configuration1a(At most compatible3a) |
standard configuration1a(At most compatible3a) |
Liquid addition volume |
4–995µL,1µLstep |
4–995µL,1µLstep |
sample injection speed |
30s,96-wellplate,5ul/well |
30s,96-wellplate,5ul/well |
Other parameters | ||
Support version |
6-384Orifice plate (only supports light absorption)96Orifice plate) |
6-384Orifice plate (only supports light absorption)96Orifice plate) |
detection speed |
96Orifice plate, absorbing light6sfluorescence18s,384Orifice plate fluorescence60s |
|
Incubation oscillator |
Standard configuration, room temperature+3℃-65℃ |
Standard configuration, room temperature+3℃-65℃ |
light source |
Xenon lamp, tungsten lamp |
Xenon lamp, tungsten lamp |
adapter |
96Orifice plate absorption light adapter 96/384fluorescence/Luminous adapter 6-48Orifice plate fluorescence/Luminescent liquid separator |
|
Configuration list:
serial number |
item name |
Specifications |
quantity/unit |
1 |
Multi functional enzyme-linked immunosorbent assay (ELISA) reader |
1Taiwan |
|
2 |
Instruction Manual |
1copy |
|
3 |
power cord |
Three phase plug |
1root |
4 |
Certificate of Conformity |
1copy |
|
5 |
Warranty Card |
1copy |
|
6 |
Inspection Acceptance Form |
1copy |
|
7 |
mouse |
1a |
|
8 |
softwareUplate |
1a |
|
9 |
96orifice plate |
blackboard |
1block |
10 |
96orifice plate |
whiteboard |
1a |
11 |
96orifice plate |
Absorbing light transparent substrate |
1a |
Ordering information:
name |
Specifications |
Remarks |
market price |
Current selling price |
SGSY 1180AMulti functional microplate reading machine |
fully loaded15standard configuration4optical filter |
Excluding computers |
281000.00 |
230000.00 |
SGSY 1180BMulti functional microplate reading machine |
Full wavelength automatic grating type |
Excluding computers |
365000.00 |
300000.00 |