E-mail
13817879697@126.com
-
Phone
13817879697
-
Address
5th Floor, No. 11, Lane 4855, Guangfulin Road, Songjiang District, Shanghai
Bennett Automation Instrument (Shanghai) Co., Ltd
13817879697@126.com
13817879697
5th Floor, No. 11, Lane 4855, Guangfulin Road, Songjiang District, Shanghai
The Setra Systems 2000 series remote particle counter measures particles ranging from 0.3 μ m to 5.0 μ m with a flow rate of 1.0 CFM (28.3 LPM) or less. Can be achieved through Modbus ™ RTU or ASCII can be easily integrated into real-time monitoring or building management systems for use. As the smallest remote particle counter on the market, the 2000 series is easy to install, compliant with ISO 21501-4 standard, and comes with a 2-year warranty.
Channel size
0.3μm. 0.5μm. 1.0μm. 5.0μm
counting efficiency
50% of particles are 0.5 μ m Particles above 0.75 μ m are 100% (Composite JIS standard)
Concentration limit
﹥500, 000/ft3@5% overlap loss
light source
Long service life laser diode
zero count
﹤A Count/5 minutes (compliant with JIS B9921 standard)
traffic
0.1 or 1.0 CFM (28.3 LPM)
flow control
Critical flow orifice
Vacuum requirements
外部真空>15”(38.1cm)Hg 校准; NIST可追溯
samplingProbe/pipeline
Isokinetic sampling probe
communication interface
Modbus RTU or ASCll output (as specified in the order)
power
9-30VDC
size(L×W×H)1
1.8 '×3.7 '×1.5 '(4.5cm×9.3cm×3.8cm)
weight
0.62lb (0.28Kg)
standard
ISO 21501-4 and JIS B9921
warranty
2-year limited warranty
working conditions
50To 35 ℃ (41020% to 95% non condensing at 95 ℉
Storage conditions
200To 50 ℃ (-40Up to a maximum of 98% non condensing at 122 ℉
Optional Accessories
Constant velocity sampling probe, blow filter, data cable, vacuum and sampling pipeline.
1Excluding probe: limited to the body and installation version only;
Specifications and parameters are subject to change without prior notice.
Example of order number;SPC23014;
2000 series,0.3μmMinimum particle size.1CFM traffic, 4-channel.
|
model |
Minimum particle size |
traffic |
通道 |
||||
|
SPC2 |
2000 series |
3 |
0.3μm |
01 |
0.1CFM |
2 |
2通道 |
|
|
|
5 |
0.5μm |
10 |
1.0CFM1 |
4 |
4通道 |
1Particle size code only'5' is not applicable for particle size code '3'
2Only limited to traffic codes“01”; Not applicable for traffic code '10'