SBA-78Biosensing automatic analyzer is an intelligent instrument that uses immobilized enzymes as key components. According to the number of enzyme sensors and whether they are equipped with external dilution devices, the instrument is divided into four models, with the dual electrode type beingSBA-78B2The dual electrode dilution type isSBA-78C2The three electrode type isSBA-78B3The three electrode dilution type isSBA-78C3The dual electrode type can simultaneously measure two analytical indicators, while the three electrode type can simultaneously measure three fold indicators. Models without additional dilution function are achieved by controlling the injection volume10Multiple dilution determination, models with additional dilution function can achieve the highest250Dilute the measurement by multiple times. The instrument can analyze different objects by replacing different immobilized enzyme membranes. The entire analysis process is controlled by a microcomputer, with automatic calibration and dilution measurement. It has a wide range of applications in clinical laboratory testing, food analysis, industrial fermentation, environmental testing, and scientific research.

| Range of determination for glucose and glutamic acid | 0-10g/L(SBA-78B), or 0-200g/L(SBA-78C) |
| Lactic acid determination range | 0-5g/L(SBA-78B), or 0-100g/L(SBA-78C) |
| Measurement accuracy | The measured concentration is1glStandard glucoseCv≤2% |
| measurement time | 20second |
| Measurement cycle | Depending on the type and concentration of the sample, typically80~120In seconds. |
| sample size | 0.5ml (automatic recognition)Actually, every measurement40ul |
| sample tube | 1.5 mlor2mlDisposable centrifuge tube |
| sample tray | 15 Location, can be placed12A sample |
| Calibration interval | 1~12Sample, customizable |
| display screen | 5Inch colorLCDtouch screen |
| Determination of data memory | 2MCapacity can be recorded10When the measurement results exceed 10000, the initial data will be automatically overwritten once the records are full. |
| print output | List printing or individual printing is optional. |
| printer | built-in |
| language | Chinese or English switchable |
| safe mode | password protection |
| data interface | UDisk, output file format: comma separated files(.CSV,WindowThe default opening method isEXCEL, customizable file name. |
| Environmental temperature for use | +10 ~+35 °C |
| humidity | <90 % |
| power supply | 220±10%VAC, 50 Hz |
| power consumption | <50W |
| volume(LxBxH) | 320x 240 x 190 mm |