- Phone
-
Address
Room 1008, Building A, Vanke Vitality City, No. 31, Lane 2419, Hunan Road, Pudong New Area, Shanghai
Yanyi Scientific Instruments (Shanghai) Co., Ltd
Room 1008, Building A, Vanke Vitality City, No. 31, Lane 2419, Hunan Road, Pudong New Area, Shanghai
RoSS.FILL CGT Original Solution Packaging System
Multiple small disposable bags can be filled simultaneously. The fluid path is designed to be disposable and can accommodate various sterile connection and disconnection options.
This platform canpartPack multiple small disposable bags with capacities ranging from 1 Milliliters to1000 The milliliters vary, thereforesuitableFilling small volumes in cell and gene therapy, small volume bulk drugs, or clinical research. The option to add extra shelves makes it possible to fill larger batches.
characteristic:
• High precisionpartPack small disposable bags
• Flexible settings, with the option to connect to a second shelf
Integrated sealed clamp valve for fluid control and aseptic decoupling
• Can be filled at most simultaneously3 A disposable bag
• Easy to operate/Formula driven processing
• Reliable platform with customized disposable components
• Bag and connector independent settings
Suitable for cleanrooms and based on international biopharmaceutical standards
• Optional sampling
RoSS.FILL CGT Original Solution Packaging System
RoSS. FILL CGTGeneral Information | ||
Application field |
clinical research/Cell and gene therapy/Seed cultivation enhancement/low capacityBDS |
|
Primary packaging |
2Ddisposable bag |
|
volume/speed/quantity |
RoSS. FILL CGT 36 |
RoSS. FILL CGT 64 |
Number of bags per unit |
36 |
64 |
number of bags |
72 |
128 |
Capacity per bag |
1to1000L |
1to250mL |
Capacity (per system run) |
72L |
32L |
packagingspeed |
per hour36 x 1000mL(36L/Hour) |
each0.75hour64 x 250mL(21L/Hour) |
precisesex | ||
Accuracy of weight method |
±2 g (10 - 500 mL) / ±5 g (500 - 1000 mL) |
|
Accuracy of flow sensor |
±20 % (1 - 5 mL) / ±10 % (6 - 20 mL) |
|
reading |
1 g |
|
Weighing grade |
OIML R60 C3level |
|
Valve settings |
RoSS. FILL CGT 36 |
RoSS. FILL CGT 64 |
Quantity of each valve |
36+2 |
64+2 |
Drive Type |
motor |
|
Clamping function |
include |
|
Step function |
include |
|
Sealing function |
include |
|
Perforation function |
include |
|
Number of bubble sensors per unit |
2 |
|
Manifold setting | ||
Pump tube outer diameter |
5/8“ |
|
Main manifold outer diameter |
3/8“ |
|
Outer diameter of bag branch |
1/4“ |
|
Manifold supply |
Single Use Support + All other suppliers |
|
Dimensions and Weight |
RoSS. FILL CGT 36 |
RoSS. FILL CGT 64 |
Size (length)xwidexHigh) |
1717 x 770 x 1617 mm 67.598 x 30.315 x 63.661“ |
2317 x 770 x 1617 mm 91.220 x 30.315 x 63.661“ |
weight |
420 kg |
690 kg |
specification | ||
pump |
Watson Marlow 630 |
|
pump speed |
7200 mL/min (3/8“ ID) |
|
flow sensor |
upgrade |
|
Sample scale |
include |
|
Weight limit of sample scale |
6000 g |
|
Readability of sample scale |
1 g |
|
Pressure sensor socket |
2 |
|
Label Printer |
include |
|
Filter configuration |
include |
|
sizeCU(Long)xwidexHigh) |
1294 x 902 x 1705 mm 50.945 x 35.512 x 67.126” |
|
other | ||
Design according to the following requirements |
ASME/ CE/ UL/ASME-BPE/cGMP/ 21 CFR Part 11 |
|
power supply |
230V 50Hz/ 115V 60Hz |
|
can be used asI/O |
is |
|
Control system manufacturer |
Siemens S7 / Delta V/ PCS 7 |
|
communication |
OPC UA interface available |
|
Mobile system |
is |
|