1 PY-DP8060 Product Overview
RoHSIt is EU legislation restricting access to electronic and electrical equipment(EEE)The full name of the regulation on the use of toxic and harmful substances in China is the "Directive on the Restriction of the Use of Certain Hazardous Substances in Electronic and Electrical Equipment"(Restriction of Hazardous Substances)This directive is mainly used to standardize the material and process standards of electronic and electrical products, making them more conducive to human health and environmental protection. Initially, the directive specified lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers in electronic and electrical products6The usage limit of the substance.2015year6moon4The European Union Gazette(OJ)ReleaseRoHS2.0Revision instruction(EU)2015/863, officially willDEHPTheBBPTheDBPTheDIBPListed in the restricted substance list. After the issuance of this revised directive, EU member states are required to2016year12moon31Recently, this directive has been translated into regulations of various countries and implemented. and2019year7moon22From now on, all electronic and electrical products exported to Europe(Except for medical and monitoring equipment)All must meet this restriction requirement;2021year7moon22From now on, medical equipment(Including extracorporeal medical equipment)And monitoring equipment(Including industrial monitoring equipment)It will also be included in the scope of this control.
To test electronic and electrical products4The content of phthalates in the electronic sample needs to be crushed, extracted, purified, concentrated, and finally detected using corresponding instruments. The pre-treatment of the sample in the early stage requires a long time.This method is used to analyze the regulation of electronic and electrical productsRoHScommand(IEC 62321 - 8:2017)Restricted phthalates.Py/TD-GCasPy/TD-GC-MSThe low configuration alternative solution has been extensively tested and can basically meet the requirementsRoHS 2.0Screening requirements for ortho benzene.
Compared to traditional Soxhlet extraction, usingPy/TD-GCThe combined system can quickly screen for phthalates in electronic and electrical products. This method does not require pre-treatment and eliminates organic solvents. The analysis results can be obtained by directly weighing the sample and analyzing it on the machine, which is simpler and faster.
2 PY-DP8060Main applications
RoHS 2.0Designated polymerinfour itemsPhthalate esters
3 PY-DP8060Functional Features
Quick screening, easy to operate;
The temperature control of thermal cracking adopts programmed heating design;
Thermal cracking is equipped with a heat dissipation system, and the instrument preparation time is short;
The injection process has no cold spots, ensuring repeatability and accuracy;
Sample preparation does not require organic solvents;
Suitable for various sample matrices (liquid, solid);
Professional software, easy to operate;

5 Analysis and Testing
5.1 Analysis Process
(1) Sample processing
The total mass of the weighed sample is:0.5-1.0 mgFix the sample in the middle of the quartz tube with tweezers and secure both ends with quartz wool; Add the liquid sample directly onto the quartz wool. Then put the quartz tube into the heating device.
(2) Reference conditions
Cracking program temperature rise:100℃(72℃/min)280℃(2 min)
Transmission line temperature:300℃
Sample diversion injection:solidEntersampleThe weight needs to be within0.5mgto1mgBetween them, the liquid should be1 ul左右
FIDDetector temperature:300 ℃, injection port temperature:300 ℃
temperature-programmed 100 ℃ (maintain)1 Minute by minute20 ℃ rises to300 ℃ (maintain)14Minutes)
(3) Reference results
Reference retention time and time window for each component of phthalate ester compounds

5. 2 Standard curve and linear range
Configurationcontentfor50 ppmThe100 ppmThe250 ppmThe500 ppmand1000 ppmofDBPStandard solution,withThe peak area of quantitative components is plotted on the y-axis,Concentration on the horizontal axis,Draw standard curve.
The standard curve and linear correlation coefficient are shown below:


6 Work environment
Environmental Requirements
The operating temperature is15°C~35°C, using temperature18°C~28°C,1.5hThe internal temperature change is not greater than2°CDue to the large temperature difference in the environment, air conditioning should be installed.
The relative humidity must be within≤75%Can be equipped with a dehumidifier.
Operation and storage require a non condensing and non corrosive environment.
The maximum dust particle density is80mg/m3.
Electrical Requirements
Voltage range:220V±10VThe frequency is50Hz.
The power supply has a grounding wire; No strong electromagnetic interference.
have10AInsert board for three pin plug3One.
If funds permit, it is recommended to install an uninterruptible power supply(UPS)Power requirements10KVAThe above.
Requirements for instrument installation test bench
Provide sufficient laboratory space for installationPYTheGCThePCDevices such as printers.
The experimental platform must have a flat surface and be stable enough to support the weight of the entire system, includingPYTheGCThePCEquipment such as printers, with a total weight of approximately120kg.
Leave no less than around the experimental platform0.5mPlease do not place it against the wall due to the wide operating and maintenance space.
Suggest installing exhaust equipment above the experimental platform.
Suggested size for instrument installation laboratory bench: length250cm×wide80cm×tall75cm.
Requirements for sample processing test bench
The sample processing experimental platform is mainly used to place electronic scales and perform sample processing.
The sample processing experimental platform should be stable and durable, and reduce the generation of vibration.
Electronic scales should not be placed near ventilation openings or fans, as strong airflow can affect the weighing results.
The sample processing experimental platform should also have sufficient space for sample processing.
Suggested size for sample processing experimental bench: length120cm×wide80cm×tall75cm.
If space is limited, two long instrument installation and sample processing test benches can also be used150cmThe experimental table.