E-mail
radonbj@126.com
- Phone
-
Address
No. 36, Majiabao West Road, Fengtai District, Beijing
Beijing Leiteng Environmental Technology Co., Ltd
radonbj@126.com
No. 36, Majiabao West Road, Fengtai District, Beijing
One of the important equipment of the neutron radiation monitoring system in nuclear radiation areas is the neutron detection device, which is mainly used to monitor the neutron dose level in areas with activated neutrons. The monitoring information is converted into digital signals and uploaded to the local processing box through paired 4-core shielded cables. The regional neutron detection device communicates with the local host using RS485 protocol.
Technical Indicators:
Detection device type: 3He proportional counter tube
Detecting radiation: neutrons
Measurement range: 0.01 μ Sv/h to 100000 μ Sv/h
Communication between RS485 protocol and local processing box
Relative inherent error: ≤ 20%
Energy response: The energy response deviation for Am Be and 252Cf sources is less than 50%
Alarm response time ≤ 10s
Equipment enclosure protection level IP65
Reliability: MTBF (Mean Time Between Failures) ≥ 5000h
Maintainability: MTTR (Mean Time to Repair) ≤ 0.5h
According to the method specified in GJB152A, the electromagnetic compatibility has passed 11 experiments.
Dimensions: Φ × H ≤ 200X260mm
Weight: ≤ 4kg
Note: The size and weight can be customized according to application requirements.
The device can work normally under the following environmental conditions:
Environmental temperature: 0 ℃~60 ℃
Humidity: 95% (temperature of 45 ℃), with condensation
Salt spray, oil spray, mold
Shock and vibration
Performance characteristics:
The equipment adopts waterproof and dustproof design
The power circuit of the detection device adopts overvoltage protection, overcurrent protection, and surge protection measures
The detection device adopts a sealed aluminum shell to ensure the integrity of internal electromagnetic shielding
Intelligent probe with self checking function and digital signal communication
Anti impact, anti vibration, and meet the electromagnetic compatibility requirements of GJB152A
General configuration of regional neutron monitoring instrument:
Processing display unit RAM-300, 2 units;
3He管neutron detectorDT-2, 6;
Lenovo computer, 1 unit;
Data display software, 1 set;
Connect power cables and signal lines, 1 set.