Welcome Customer !

Membership

Help

Kunshan Sainikos Automation Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Redox transmitter

NegotiableUpdate on 05/18
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Product features: ■ 144X144 microcomputer designed pH/redox transmitter ■ Large LCD display screen with operation status and special symbol display ■ Equipped with backlight sensor, automatic and manual backlight setting function ■ Provided with electrode automatic cleaning device contact, program setting output ■ Provided with solution grounding function, can eliminate sample electrification interference ■ Equipped with automatic calibration; Manual/automatic temperature compensation function and non periodic calibration warning ■ Password lock and combination key for setting and calibration mode design, enhancing safety protection ■ Equipped with two 0/4~20mA outputs corresponding to pH/ORP settings Temp.

Product Details

product details

Typical applications:Widely used in wastewater treatment, purified water, circulating water, boiler water and other systems, as well as electronics, electroplating, printing and dyeing, chemistry

In the fields of food, pharmaceuticals, and other manufacturing processes, it has demonstrated outstanding performance in applications such as large-scale wastewater treatment plants and industrial process monitoring.

Technical Parameter

model

PH-2100

range

pH: -2.00~16.00pH

ORP: -1999~+1999mV

Temp.:–30.0~130.0℃

accuracy

pH: 0.01pH±1Digit

ORP: 1mV±1Digi

Temp.:0.1℃±1Digi

temperature compensation

Automatic temperature compensation Pt1000, NTC30K dual-mode selection, manual temperature compensation, key setting

Analog Output

Two isolated 0/4-20mA outputs, programmable for pH/ORP and temperature settings

Digital Output

not have

Relay

High and low points, hysteresis can be set by any program, with two sets of ON/OFF outputs,

Clean the contact points

ON 0-999 seconds/OFF 0-999 hours, time program setting controls a set of ON/OFF outputs

Input impedance

>1012Ω

ambient temperature

-10~50℃

power input

88-265VAC wide power supply, 50/60Hz

Instrument size

144mm×144mm×115mm(H×W×D)

Hole size

138mm×138mm (H×W)

Installation method

Wall installation, panel installation, and circular pipe hanging installation

protection grade

IP65

weight

0.8Kg