E-mail
790895870@qq.com
-
Phone
13916481105
-
Address
W-12, 2nd Floor, No. 18, Lane 945, Changta Road, Shihudang Town, Songjiang District, Shanghai
Shanghai Anyi Scientific Instrument Co., Ltd
790895870@qq.com
13916481105
W-12, 2nd Floor, No. 18, Lane 945, Changta Road, Shihudang Town, Songjiang District, Shanghai
Our company is a manufacturer and supports customization
GTT/STT series heat flow sensor/radiation heat flow sensor
Heat flow sensors/radiation heat flow sensors are used in ground and flight aviation tests, heat transfer research, flammability testing heat flow standards, fire testing, material development, and kiln development.
Sensor Description
structure
Accurate, sturdy, and reliable sensors.
Oxygen free copper body ensures long lifespan and signal stability of the sensor.
Equipped with flanges for easy installation. Attachment
Optional configurations: including four installation configurations, detachable windows, water cooling, and conduction cooling. When the working temperature exceeds 200 ° C, water cooling should be chosen.
Window attachments (optional window materials such as sapphire, quartz, zinc sulfide, zinc selenide, etc.), measure radiation heat flux (total radiation, infrared radiation).
The standard installation configuration is shown in the figure. Including smooth surfaces with flanges; Threaded surface with flange; Smooth surface without flanges; Threaded surface, without flange
Sensor parameters:
Type: Gardon style/Schmidt Boelter style
Optional ranges: 30000, 20000, 15000, 10000, 5000, 4000, 2000, 1000, 500, 300, 250, 200, 150, 100, 50, 25, 20, 10, 5, 2, 1 (kW/m2)
Output signal: Millivolt linear output (with correction factor)
Sensor absorption rate: greater than 0.95
High tolerance sensor temperature: 200 ° C (if it exceeds 200 ° C, cooling water is required, and the temperature tolerance can reach 3000 ° C)
Repeatability: ± 0.5%
Calibration extended uncertainty: ± 3%, including factor k=2 (for approximately 95% confidence level) (± 5% for 5000kw/m2 and above)
Non linearity: ± 2%
Wire: Copper double stranded wire, Teflon insulation, copper wire braided shielding, 1m standard length, bare end
Size: 25 (diameter) x 25 (length) mm; flange diameter 45mm (please contact us for multiple sizes and shapes)
Optional test outputs: 1. Data acquisition host DQ-100; 2. Amplifier TFT series
Radiation heat flux sensor window material wavelength range window material (1mm, unless otherwise specified) symbol (insert model) wavelength range (um) sapphire (0.5mm) S 0.2-5.5 quartz (0.5mm) QW 0.12-4 calcium fluoride CaF2W 0.3-11.5 KRS-5 KRS-5W 0.6-50 zinc sulfide ZnSW 0.5-14.5 barium fluoride BaF2W 0.5-12.5 zinc selenide ZnSeW 0.5-22 cadmium telluride CdTeW 0.8-30
Sensor customization
Type probe diameter (mm) (standard size 25mm)
Range (kW/m2)
Window water-cooled flange thread
SchmidtBoelter
Gardon Y N Y N Y N Y N
STT GTT -XX -YY -R -W -F -T
Example: GTT-25-100-RWF refers to a Gardon type sensor with a diameter of 25mm, a range of 100kW/m2, a viewing window (only measuring radiant heat flux), water cooling, a flange, a smooth surface (without threads), and a signal line length of~1.8m (standard length)
TTF series amplifier
TTF series transmitter input/output/power three-way full isolation, anti-interference, and input/output selection range
Wide circumference, high accuracy, and easy installation of guide rails for inspection and maintenance. Applied to data acquisition, signal transmission conversion, and DCS distributed control
Manufacturing system.
Product Features:
Compact size, beautiful appearance, integrated isolation and transmission
Good linearity, good temperature characteristics, high conversion accuracy, stable performance
Equipped with power indicator light, signal fault and range indicator light
Standard DIN35 rail installation, easy to install and disassemble, suitable for dense installation
Fully isolated between input/output/power supply, with strong anti-interference ability
Product Parameters:
输入: 0-10mV 0-15mV 0~20mV 0-100mv 0-200mv
Output: 0-5V, 1-5V, 0-10V
Power supply voltage: 24V DC+10%
Accuracy: 0.2%
Working temperature: -20 ℃~70 ℃