Radiation food and drug rapid detection device

NegotiableUpdate on 02/11
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The PPSL-18 rapid detection instrument for irradiated food and drugs is a light stimulated luminescence system based on the principle of photoluminescence (PSL). It consists of a pulse excitation light source (infrared excitation light source), a sample chamber, a detection probe, a photon counting system, and a control unit. It is used to quickly identify whether food and drugs have been irradiated or not, in accordance with the 2025 version of the national drug standard "Guiding Principles for Photoluminescent Identification of Irradiated Traditional Chinese Medicine"
Product Details

PPSL-18Radiation food and drug rapid detection devicebelong toLight stimulated luminescence(PSL systemalso known asLuminescent system,Composed of a pulse excitation light source (infrared excitation light source), a sample chamber, a detection probe, a photon counting system, and a control unit, it is used for rapid identification of whether food and drugs have been irradiated. Widely used for rapid screening of traditional Chinese medicine decoction pieces, solid preparations containing raw powder, spices, seasonings, etc., it can effectively identify samples treated with ionizing radiation (irradiation) at a dose of 1kGy or above. Comply with the Guiding Principles for Photoluminescent Identification of Irradiated Traditional Chinese Medicine in 2025 National Drug Standard "Guiding Principles for Photoluminescent Identification of Irradiated Traditional Chinese Medicine"

As an efficient and economical method for microbial sterilization, inhibition of plant germination, and extension of food shelf life, irradiation technology has been widely used in fields such as food and medicine. However, just like genetically modified foods, the effects of irradiated food and drugs on the human body are still unknown. For the long-term consideration of human health and safety, developed countries such as Europe and America have long strengthened their supervision of irradiated food and drugs. In China, with the increasing public awareness of food safety and the government's attention, food and drug regulatory authorities have also begun to strengthen their supervision of irradiated food and drugs to prevent irreversible consequences caused by the abuse of irradiation technology in the future.

Chinese medicinal herbs, spices, etc. usually contain or carry minerals (inorganic substances, silicates, etc.). When exposed to ionizing radiation (irradiation), these minerals store energy through their structural voids. When stimulated by excitation light, these stored energies will be released in the form of photons.OslIdentification method is a method of identifying whether traditional Chinese medicine decoction pieces and solid preparations containing raw powder have been subjected to ionizing radiation (irradiation) based on the number of photons released.

PPSL-18Radiation food and drug rapid detection deviceadoptdiameter50mmPi'sSample dishThe required sample quantity is extremely small,Single sample testing time only takes 60 seconds.

The PPSL-18 is designed to be compact and small in size, requiring only one data cable to connect to a computer for operation. All testing processes, except for sample collection and placement, are completed by software operation.

Product Introduction

Adopting the principle of photoluminescence (PSL), it complies with the EU standard EN13751 and equivalent national standards.

Simultaneously display and record the photon count per second and photon release curve graph.

The combination of photon counting and photon release curve improves the recognition rate of suspected samples.

High sensitivity can identify low-dose radiation below 1kGy.

Extremely low sample pre-treatment requirements; No need for any chemical reagents.

The duration of a single sample testing is60Seconds, especially suitable for rapid screening.

Automatic data saving, data callback, and report generation for individual and batch samples.


Technical Specifications

Measurement principle: Light stimulated luminescence(PSL)The law

Excitation light source: infrared arrayLED

Photon Counting Detector: End Window Photomultiplier Tube Integrated High Voltage Power Supply

Dark count rate: typical value of 60CPS, maximum value of 150CPS (after avoiding light for 30 minutes at room temperature of 25 ℃)

Standard threshold: 700/5000

Preset testing cycle: 60 seconds

Sample dish: 50mm diameter Petri sample dish, made of aluminum alloy material, durable and long-lasting

Power supply: 220VAC/50Hz, power: 12W

Dimensions: Dimensions: 210 (W) X 260 (H) X 330 (D) mm

Weight: 11.2kg (net weight)