The RU series UV visible spectrophotometer has advanced functions such as quantitative measurement, wavelength scanning, time scanning, multi wavelength testing, DNA protein testing, etc. It has the characteristics of high precision, high resolution, low noise, and low stray light, and is particularly suitable for fields such as scientific research, pharmaceutical inspection, life sciences, environmental monitoring, metallurgy and chemical engineering.
RU series UV visible spectrophotometerfeature
The instrument has low drift and high stability, can compensate for reference changes, and is suitable for long-term testing
◆ Push pull sample cover design, spacious sample chamber, can accommodate 5-100mm colorimetric dishes, supports various accessory accessories required for testing
High quality optical and electronic components ensure high precision and low stray light of the instrument
◆ Large screen dot matrix LCD, clear display of graph data, user-friendly interface, easy operation
◆ Smaller spectral bandwidth, higher resolution, meeting pharmacopoeia requirements, and wider applicability
◆ Long life environmentally friendly flange seat deuterium lamp, no optical debugging required for lamp replacement
Real time saving of graphs and data, with the ability to maintain power off, printable, and exportable to PC
◆ Sturdy base plate and optical bracket, imported key components, durable and long-lasting
◆ It can be operated offline or connected to a computer, making online operation more convenient and expanding data processing and storage capabilities
RU series UV visible spectrophotometerSpecifications
spectral bandwidth |
1.8nm∕1.8nm /0.5、1、1.5、2、4、5nm 可调 |
wavelength range |
190~1100nm |
optical system |
Single beam/quasi dual beam/dual beam optical path |
Wavelength accuracy |
±0.1nm(656.1nm); ± 0.3nm (full area) |
Wavelength repeatability |
≤0.1nm |
Wavelength setting |
Automatic wavelength setting, wavelength resolution of 0.1nm |
Luminous range |
0~200%T, -4~4A, 0~9999C, 0~9999F |
Photometric accuracy |
±0.002A (0~0.5A), ±0.004A (0.5~1A), ±0.2%T (0~100%T) |
Photometric repeatability |
≤0.001A (0~0.5A), ≤0.002A (0.5~1A), ≤0.1%T (0~100%T) |
stray light |
≤0.03%T (220/360nm) |
Baseline drift (stability) |
≤0.0004A/h (500nm,0A) |
noise |
± 0.1% T (0% T line); ± 0.3% T (100% T-line); ± 0.0004A (at 500nm) |
baseline flatness |
±0.001A |
scanning speed |
Three adjustable speeds: high, medium, and low |
detector |
Imported silicon photodiode |
light source |
Imported long-life tungsten and deuterium lamps (without optical debugging) |
display |
3-inch 128 * 64 dot matrix LCD display/6-inch 320x240 dot matrix LCD display |
keyboard |
Digital touch film button |
data interface |
USB port |
Print output |
Micro printer; PC printer (for online use) |
PC software |
PC scanning analysis software (standard) |
power supply |
AC90-250V, 50/60Hz |
Overall dimensions |
430x320x195mm∕635x515x255mm |
weight |
12kg∕26 kg |