- Phone
-
Address
No. 11 Yuehe Road, Chengyang District, Qingdao City
Qingdao Zhongrui Intelligent Instrument Co., Ltd
No. 11 Yuehe Road, Chengyang District, Qingdao City

Aerosol photometer cleanroom detection, high-efficiency filter leak detection
ZR-6012 typeAerosol photometer cleanroom detection, high-efficiency filter leak detectionIt is a specialized detection device designed based on the Mie scattering principle for detecting leaks in high-efficiency filters. Based on upstream detection concentration and scanning detection of downstream concentration, the leakage rate of high-efficiency filters can be obtained in real time. Suitable for on-site testing of high-efficiency filter systems by third-party testing companies, pharmaceutical companies, metrological verification units, etc.
1. Third level user management and audit tracking function to ensure data integrity;
2. Wide range of aerosol concentration detection;
3. High definition color touch screen, convenient interaction;
4. Multiple types of aerosols such as DOP and PAO can be set up;
5. Large capacity data storage, real-time saving of sampled data
6. Data can be exported through a USB flash drive or printed using a built-in printer;
7. Optional built-in battery for easy operation without power supply;
8. Small in size and light in weight, combined with specialized packaging luggage, convenient to carry;
9. It has multiple user-friendly functions such as sound and light alarm when exceeding the set value, automatic prompt for replacement of high-efficiency filters, and automatic protection for fault detection.
GB/T 16157-1996 Determination of particulate matter and sampling methods for gaseous pollutants in exhaust gas from fixed pollution sources
GB/T 11605-2005 Humidity Measurement Method
HJ/T 48-1999 Technical Conditions for Smoke and Dust Samplers
JJF 1800 - 2020aerosol photometerCalibration standards
2010 Drug GMP Guidelines - Plant and Equipment
ISO 14644-3:2005 Cleanrooms and Related Controlled Environments - Part 3: Test Methods
GB 50073-2013 Code for Design of Cleanrooms
GB 50591-2010 Code for Construction and Acceptance of Cleanrooms
YY 0569-2005 Biosafety Cabinet
JJF 1815-2020 Calibration Specification for Biosafety Cabinets
| Main parameters | parameter range |
| Testing medium | PAO, DOP and other types of media |
| Size (W × D × H) | 300*330*184 |
| Total weight | About 8Kg (including battery) |
| Gross weight including packaging | About 14Kg (outer packaging, main unit, accessories, etc.) |
| Power supply | Input AC220V ± 10%, 50Hz |
| Total power consumption | <200W |