Price of WKL-202 microcoulometric sulfur analyzer
WKL-202 microcoulometric sulfur analyzerOn the basis of retaining the advantages of the prototype product in chemical analysis, significant improvements have been made to the host's sampler, stirrer, etc. Particularly noteworthy is the new design of a microcoulomb automatic titration system, which is controlled by a microprocessor to control the entire titration process, including automatic stabilization of the titration cell potential, control of the sampler, control of titration current, and integration and display of titration electricity. The temperature of the three-stage converter is also controlled by a microprocessor, with a temperature control accuracy of ± 10 ° C. In order to meet the needs of low concentration gas determination, an online automatic gas injection system has been specially designed and manufactured. After the modification, the automation level and high-tech content of the instrument have been improved, and the appearance of the instrument has also been greatly improved.
The sample is first subjected to oxygen combustion in a high-temperature conversion furnace to convert all the sulfur contained in the sample into SO2, and then absorbed through a titration cell. The electrolyte in the titration cell is potassium iodide acetic acid solution. The instrument uses iodine produced by electrolysis to perform microcoulometric titration on SO2 entering the titration cell. Use the potentiometric method to control the endpoint of titration, and calculate the amount of electricity consumed throughout the titration process to determine the sulfur content of this injection.
Implementing automatic titration can eliminate human errors caused by unskilled operators. As long as the 'Start' button is pressed, the microprocessor sends instructions to the sampler to gradually inject the sample into the conversion furnace. As SO2 in the sample enters the titration cell, the potential of the titration cell decreases. The single-chip microprocessor controls the titration current based on the amount of potential decrease. Until all SO2 entering is consumed and the potential of the titration cell returns to its initial value before titration, the titration will automatically stop and display the total sulfur content ng of the sample in this injection.
The temperature control system of the three-stage converter can easily set the temperature of each section of the conversion tube and view the actual temperature of each section. It also adds a thermocouple cold end compensation link to ensure temperature control accuracy. When there is an over temperature or thermocouple breakage, the instrument will automatically alarm.
After the improvement of the mixer and sampler, the structure is more reasonable, the use is more convenient, the overall appearance of the instrument is more compact and beautiful, and the overall integrity is stronger.
WKL-202 microcoulometric sulfur analyzerThree items have been applied for: appearance, isolation electrolysis power supply, and sampler control system.
WKL-202 microcoulometric sulfur analyzerTechnical indicators:
1. Sulfur measurement range: Liquid sample: 0.1~1000mg/l
Gas sample: 0.1~1000ppm
Zui small inspection measurement: 0.1ppm
2. Repeatability: ± 2%
3. Power supply voltage: 220V ± 22V A.C. Power supply frequency: 50Hz ± 0.5Hz
4. Environmental temperature: 5 ℃~35 ℃
5. Relative humidity of the environment: ≤ 85%
6. Dimensions: Length x Width x Height=740 x 400 x 374 (mm)
7. Instrument weight: 25 kilograms