Power frequency voltage breakdown tester

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The SDJ-100KV power frequency voltage breakdown tester is controlled by a computer, and the test curve can be observed online during the test process; Automatically store test conditions and test results data, and can access, display, and print them.

Product Details

Applicable materials:

SDJ-100KVPower frequency voltage breakdown testerMainly suitable for testing the breakdown strength and withstand time of solid insulation materials such as plastics, films, resins, mica, ceramics, glass, insulation oil, insulation paint, cardboard and other media under power frequency voltage or DC voltage; The voltage breakdown tester is controlled by a computer, and the test curve can be observed online during the test process; Automatically store test conditions and test results data, and can access, display, and print them.


Basic functions:

The voltage breakdown tester is controlled by a computer, and the data acquisition method is through photoelectric isolation, effectively solving the anti-interference problem during the test process. The software is easy to operate and can display dynamic curves in real time. At the same time, the boost rate is infinitely adjustable, and the boost rate can be adjusted according to one's own needs, making the boost rate truly uniform and accurate, and accurately testing the leakage current data. Real time drawing of experimental curves, display of experimental data, accurate judgment, and the ability to save and analyze. Print the test data report. And it can automatically determine the breakdown of the sample and collect breakdown voltage data and leakage current. At the same time, it can quickly reduce the voltage and automatically reset to zero at the moment of breakdown. The software system is easy to operate, has stable performance, and is safe and reliable.


Structural features:

1. Mainly composed of: boosting system (high-voltage transformer), Siemens PLC control system, electrodes, oil tank, electrode positioning frame, computer data processing system, software, etc;

2. Computer - Siemens PLC Control System - Voltage Regulating Device - Step up Transformer - Sample;

3. High voltage transformers mainly generate the DC voltage required for the sample;

4. A voltage regulator is used to regulate the input voltage of a step-up transformer to generate the required input voltage for high voltage;

5. Voltage measurement is mainly measured from the measuring end of the high-voltage transformer, which is linear with the measuring end of the high-voltage transformer;

6. The discharge system will automatically discharge after the experiment is completed to avoid the harm of discharge to personnel.


SDJ-100KVPower frequency voltage breakdown testerTechnical parameters:

1. Related models: SDJ-10KV, SDJ-15KV, SDJ-30KV, SDJ-50KV, SDJ-100KV

2. Input voltage: AC220V/50Hz

3. Output voltage: AC: 0-50kV; DC: 0-50kV

4. Output power: 5kVA

5. Measurement range: AC: 0-50kV; DC: 0-50kV

6. High voltage classification: 0-5KV; 0-10KV; 0-15KV;0-20KV

7. Measurement error: ≤ 2%

8. Boosting rate: 5V/S --5kV/s

9. Voltage endurance time: 0-6H (can be set through software)

10. Voltage accuracy: ≤ 1%

11. Test method:

DC test: 1. Breakdown test 2. Voltage withstand test 3. Gradient test

Communication test: 1. Breakdown test 2. Voltage withstand test 3. Gradient test

12. Breakdown detection stop method: voltage detection stop, current detection stop

13. Electrode specifications: Two pieces of ¢ 25mm, one piece of ¢ 75mm

14. Host size: 1000 * 750 * 1350mm (length * width * height)

15. Whole machine weight: 150kg

16. Level 10 security protection:

(1) Threshold protection

(2) Voltage zeroing protection

(3) Terminate voltage protection

(4) Software misoperation protection

(5) Overpressure protection

(6) Test short-circuit protection

(7) Leakage protection

(8) Independent grounding protection

(9) The experiment ends with fully automatic discharge protection. After discharge, the software displays the residual power, making it convenient for the experimenter to observe the voltage of the high-voltage column head

(10) When the experimenter is not discharging, the box door switch is always closed, and only opens when the discharge is completed