- Phone
-
Address
New City Science and Technology Industrial Park, Xincheng District, Xi'an City
Shaanxi Fusaisi Instrument Co., Ltd
New City Science and Technology Industrial Park, Xincheng District, Xi'an City
p-LLA4160_OCAPortable infrared oil concentration analyzer (infrared oil analyzer)Principle Introduction:
p-LLA4160'OCA is a professional organic component infrared spectroscopy absorption analyzer designed based on the principle of photometry. It identifies types by scanning spectra and tests concentration through transmittance. Perform rapid analysis on extraction.
P-LLA4160'OCA Auto is a fully automatic water oil analyzer. Simply insert the sampling tube into the water sample, press the<Analysis>or<Continuous Analysis>button, and the system will automatically extract and analyze the water sample. Support direct analysis, which means analyzing externally prepared extraction solutions.
This instrument is a fixed purpose analysis system designed for the analysis of oil concentration in water bodies. Suitable for monitoring the content of petroleum (mineral oil), animal and vegetable oils, and total oil in various water bodies and soils such as surface water, groundwater, seawater, domestic water, and industrial wastewater. It can also be used for standard sampling methods to support the monitoring of oil content in flue gas (cooking fumes in the catering industry). It can also be used for purity detection of organic reagents and measurement of the total amount and amount of organic compounds containing various C-H bonds.
Support standard methods:
Determination of petroleum and animal and vegetable oils - Infrared photometric method - GB/T 16488-1996
Infrared oil analyzer function:
Built in microcontroller microcomputer
Quickly detect parameters and temperature values, and perform temperature correction and cross correction
100 sets of data records, can be set to automatic or manual recording
RS232/485 duplex interface, capable of online sampling with microcomputers
Software zeroing, calibration of standard samples or substitutes
All operation keyboard settings, window prompts
Application of water oil detector:On site or laboratory water oil analysis
Application of portable oil analyzer:
Environmental Monitoring
Pollution source detection
On site analysis of industrial water quality
Scientific research laboratory analysis
Tester function:
On site LCD 4 × 16 character rotating display of multiple environmental parameters
Over limit alarm, alarm limit can be set
RS232/RS485 communication interface supports serial communication and can be connected to a computer
100 sets of data records. Readable, output or print
Pump sampling and sampling
Users can also calibrate or calibrate themselves
Portable infrared oil concentration analyzer (infrared oil analyzer)technical indicators:
BD7 host test ADC resolution: 1/60000;
Accuracy: ± 1-2% reading (general);
Long term stability:+/-10%/year (generally);
Analyzer response time:<10ms;
Probe response time:<1min;
Instrument usage environment: Temperature: -10 ℃ to 60 ℃; Humidity: 10% to 90% R (no condensation);
Instrument storage environment: temperature: 0 ℃~50 ℃; Humidity: 10% to 80% R (no condensation);
Electrochemical probe direct sampling: temperature: 0-40 ℃; Pressure:<1.1 kgf/cm2;
Power supply: 9V rechargeable battery with 220V AC charger:
Continuous usage time:>24 hours per charge; The battery should be stored for one week.
Detector technical parameters:
Scanning range: 3.0-4.1um; supports 2-band or 3-band analysis
Resolution: 0.1um;
Accuracy: ± 1% reading or ± 1% of full scale, whichever is greater
Dynamic range: 1:100/1000;
Range: LDLx (100~1000);
Electrical function and performance:
Please refer toBD4/BD5Smart transmitter/Introduction to Measurement and Control Devices;
Chassis packaging:NEMA 1 /IP10; NEMA 4 / IP56;
Chassis dimensions: W240*H132* D240/360;
Instrument quality assurance: 1 year;
Specification and model of portable infrared oil analyzer
Specification Model |
Optical range |
Testeable range |
||||
pLLA4160-TF3041-OCA |
0.02-2.0mg/L |
0.002-1000mg/L |
||||
pLLA4160-TF3041-OCA-Auto |
0.02-2.0mg/L |
0.002-20mg/L(Automatic Extraction) |
||||
20-1000 mg/L(Artificial extraction) |