Overview
The portable trace water analyzer MT 1000 is a trace water analyzer specially developed by our company to measure the moisture content of chlorine gas and hydrogen chloride for the production needs of chlor alkali chemical industry and downstream industries. The MT1000 uses traditional electrolysis principles to measure the moisture content in gases and is compatible with the German company CMC's 100 liter/hour flow rate micro water sensor. It has the characteristics of compact design, high sensitivity, accurate measurement, fast response, durability, and extremely low maintenance. The cleaning and regeneration of the sensor film layer is simple, and users can easily complete it. The electrolysis method is an absolute humidity measurement method, and the analyzer does not need to be recalibrated.
Product Details

Portable electrolytic micro water analyzerMT 1000Overview

Our company has developed a portable trace water analyzer MT 1000 specifically for measuring the moisture content of chlorine gas and hydrogen chloride in response to the production needs of chlor alkali chemical industry and its downstream industries. MT1000 adopts the traditional electrolysis principle to measure the moisture content in gas, and is compatible with the original German CMC 100 liter/hour flow rate micro water sensor. It has the characteristics of compact design, high sensitivity, accurate measurement, fast response, durability, and extremely low maintenance.

The portable micro water analyzer MT 1000 micro water sensor is a non consumable component, and the sensor membrane is easy to clean and regenerate. Users can easily complete it themselves. The electrolysis method is an absolute humidity measurement method, and the analyzer does not need to be recalibrated.


Portable electrolytic micro water analyzerMT 1000 application areas

Chlor alkali industry

petrochemical

Polycrystalline silicon industry

Gas/heat treatment inside the furnace

Industrial gases

semiconductor industry

pharmaceutical industry


Portable electrolytic micro water analyzerMT1000 measurement principle

The electrolysis method micro water sensor utilizes the principle of electrolyzing water molecules into hydrogen and oxygen. This sensor consists of a glass cylinder and two parallel electrodes. The electrode material is selected according to the specific application (usually made of platinum wire), and a thin layer of H3PO4 phosphate film is coated between the two electrodes. The electrolysis current that occurs between the two electrodes decomposes the water in the acid into H2 and O2. The final product of this process is P2O5, which is a highly hygroscopic substance. Therefore, water is absorbed from the sample gas, and through continuous electrolysis, a balance is established between the water content of the sample gas and the water content after electrolysis. The electrolysis current is proportional to the water content of the sample gas, and the signal is amplified and displayed. Data readout. This principle is used to measure Xe、Ar、Kr、He、D2、F2、N2、H2、O2、O3、HBr、PH3、SF6、Freon、C2H2、CO2、CH4、Natural gas, Especially suitable for measuring trace moisture in high-purity acidic gases such as Cl2, HCl, SO2, H2S, and many other gases (except for a few gases that undergo chemical reactions with phosphoric acid).


Technical parameters of portable electrolytic micro water analysis MT 1000 instrument

Analyzer

Model: MT 1000

Range: 0~10/100/1000/2500 ppm (user customizable range)

Range mode: automatic, manual

Accuracy: ± 1% of the full range

Resolution: 0.1 ppm

Display: Touch screen

Alarm output 2 alarm relays

Analog output: 4-20mA

Power supply: 220VAC/50~60Hz, built-in rechargeable lithium battery

Power consumption: 8W

Environmental temperature: -10 ℃~+50 ℃

Shell: portable, protection level IP65

Dimensions and Weight: Width 340mm, Height 290mm, Depth 220mm; Weight Approximately 3 Kg

Sensors

Model: TMS-H/P/100

Measurement poolHastelloy

Electrode Material: Platinum

Sample gas flow rate: 100 NL/h

Sample gas pressure: 0.1 barg~0.5 barg

Sample gas temperature:+5~+150 ℃ (it is recommended not to exceed 50 ℃, otherwise cooling measures need to be taken)

Accuracy: ± 1% of the full range

Response time:<1 second

Rise time: T50<8 seconds

Detection capability: 0.1 ppm

Sealing ring: FEP

Air circuit connector: 6mm card sleeve connector

Cable: Standard 3 meters, optional up to 300 meters in length

Electrical connector: Swiss LEMO waterproof connector

Installation method: Install vertically with 3 M4 screws at the bottom

Size: Approximately 43mm in diameter and 120mm in height; Weight: Approximately 1.15kg