Photovoltaic material vulcanizing instrument

NegotiableUpdate on 01/31
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The sulfurization tester for photovoltaic materials meets the standards of GB/T 16584, GB/T 33095.1, ASTM D5289, ISO6502, and DIN53529.
Product Details

Yiwei Technology's rotorless vulcanizing instrument is a new generation, reliable, accurate, and intelligent rubber rheometer. To provide good data repeatability and reproducibility for formula and quality engineers, configurable automation accessories can be used to improve testing efficiency and reduce labor costs. MES interface can be provided to facilitate the construction of factory informationization standards. A set of software operation software connects multiple instruments.

Photovoltaic material vulcanizing instrumentEV-MDR precise PT100 platinum temperature measurement, microcomputer control integrated motherboard temperature control, ultra fast temperature return speed, stable temperature 7-inch color touch screen operation, can be used independently without a computer for automatic balancing, built-in weights, automatic calibration, reducing human operation errors. The internal structure is simple and durable, and the maintenance space is more spacious. Modular design makes maintenance simple, convenient, stable and reliable, and can be used continuously with dedicated operating software. The four column suspension structure design can be used for statistical management and analysis of data, and the DSP data acquisition system is stable.

Photovoltaic material vulcanizing instrumentEV-MDR parameters:

1. Compliant with standards: GB/T 16584, GB/T 33095.1, ASTM D5289, ISO6502, DIN53529;

2. Frequency: Fixed at 1.67Hz (100cpm) or optional between 0.001Hz and 50Hz;

3. Amplitude (strain): Fixed angle ± 0.5 ° (± 7%). ± 0.2 ° (± 2.80%), ± 1.0 ° (± 14%), 3.0 ° (± 42%), or 0.001 °~90 ° (± 0.07%~± 1256%) are optional;

4. Temperature: 40~210 ℃; Higher temperature of 230 ℃ is optional;

5. Torque range: standard 0.01~225dNm, optional 0.01~50dNm;

6. Pressure range: 8300kPa (1200psi);

7. Sample volume: 4.5 cm ³;

8. Data points: Typical data ML, MH, Tan δ @ ML, Tan δ @ MH, ts1, ts2, t10, t50, t90 (optional vulcanization time), vulcanization rate, MH-ML, Sulfurization return test TR. Pressure point data PL, PH, P10, P50, P90 (optional);

9. Automated sampling system: 15 samples, 30 samples, and 120 samples are available for selection.

Repeatability and reproducibility of sulfurization tester test data

Lanxess standard rubber, testing angle ± 0.5 °, testing temperature 177 ℃, testing time 12 minutes, results of 5 parallel laboratory experiments.

test project

S'@ML

S'@MH

TC10

TC50

TC90

S''@ML

S''@MH

average

1.4

11.59

1.41

3.07

5.31

0.91

1.02

Repetitive standard deviation

0.028

0.093

0.026

0.028

0.047

0.03

0.028

Repetitive difference coefficient

2%

0.80%

1.84%

0.91%

0.89%

3.29%

2.75%

Reproducibility standard deviation

0.069

0.13

0.078

0.078

0.173

0.088

0.086

Coefficient of reproducibility difference

4.93%

1.12%

5.53%

2.90%

3.26%

9.67%

8.43%