Welcome Customer !

Membership

Help

Beijing Zhongke Xinyi Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Beijing Zhongke Xinyi Technology Co., Ltd

  • E-mail

    3038287240@qq.com

  • Phone

    18911489583

  • Address

    Changping, Beijing

Contact Now

Permanganate Index Water Quality Online Automatic Monitoring Instrument

NegotiableUpdate on 05/26
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Add a measured amount of potassium permanganate solution and sulfuric acid solution to the sample, mix and digest them. Potassium permanganate oxidizes organic pollutants in the sample, then add a measured amount of sodium oxalate solution to reduce the remaining potassium permanganate, and drip the excess sodium oxalate back with potassium permanganate solution. The titration endpoint is determined by the oxidation-reduction potential, and finally the potassium permanganate index water quality online automatic monitoring instrument automatically calculates the potassium permanganate index value.

Product Details

Principles of Measurement

Acidic potassium permanganate oxidation-reduction titration method, determination of oxidation-reduction potential endpoint:

Add a measured amount of potassium permanganate solution and sulfuric acid solution to the sample, mix and digest them. Potassium permanganate oxidizes organic pollutants in the sample, then add a measured amount of sodium oxalate solution to reduce the remaining potassium permanganate, and drip the excess sodium oxalate back with potassium permanganate solution. The titration endpoint is determined by the oxidation-reduction potential, and finally the instrument automatically calculates the permanganate index value.

Permanganate Index Water Quality Online Automatic Monitoring InstrumentMain Features

(1) The colorimetric digestion unit, injection pump precise sampling unit, and multi valve reagent switching unit are all designed with modular structures, which have lower failure rates, lower maintenance requirements, lower reagent consumption, and higher cost-effectiveness.

(2) Capable of accepting remote control commands and possessing quality control functions such as 24-hour zero drift and 24-hour range drift automatic verification, zero calibration, standard sample calibration, and standard recovery

(3) It has the function of analyzing instrument process log recording, storing raw data and operation logs for no less than 1 year.

(4) Equipped with automatic fault diagnosis function, with abnormal information recording and uploading function, such as component failure, over range alarm, over limit alarm, reagent shortage alarm and other information.

(5) Having three-level management authority

(6) After abnormal reset and power outage, the instrument automatically discharges residual reactants and returns to its initial working state. Equipped with power on alarm function, the instrument will generate alarm information when powered on again after power failure

(7) Equipped with overheat protection: If the heating temperature is too high, the heater will automatically shut down.

(8) Adopting a 7-inch touch screen and a simple operating interface, it is easy for users to learn, operate, and maintain

Typical Application

Online monitoring of permanganate index in drinking water and surface water

Permanganate Index Water Quality Online Automatic Monitoring InstrumentTechnical Specifications

measuring method

Acidic potassium permanganate redox titration method

Terminal determination

Determination of oxidation-reduction potential endpoint

Range

0-10mg/L 0-20mg/L (freely switchable)

Repeatability

Not exceeding ± 5%

Accuracy

± 5% or 0.2mg/L, whichever is greater

Resolution

0.01mg/l

Display

7-inch display screen, operation interface

sampling period

60~9999min adjustable mode, hourly measurement mode, external trigger mode optional

calibration cycle

Manual calibration and automatic calibration modes

In automatic calibration mode, it can be adjusted at any interval and any time from 1 day to 99 days

Maintenance Cycle

Usually once a month, about 30 minutes each time

Power Supply

220VAC±10%,50HZ±10%

Output

4~20mA output

Communication

MODBUS RS-232、RS-485

relay

4 channels, relay

environmental condition

Indoor temperature adjustable, recommended temperature+5-40 ℃

Humidity ≤ 85% (non condensing)

Common spare parts:ReagentsStandard solutionORP electrodeOne year maintenance package