- Phone
-
Address
B32, Science and Technology Building, 688 Zuchongzhi Road, Zhangjiang High tech Park, Pudong New Area, Shanghai
Shanghai Baige Biotechnology Development Co., Ltd
B32, Science and Technology Building, 688 Zuchongzhi Road, Zhangjiang High tech Park, Pudong New Area, Shanghai
Panasonic high-pressure sterilization potMVS-63
1Scope of application
Used for sterilization needs in hospitals, research institutions, pharmaceutical companies, and other units.
2、 Main indicators
1. Size: The cavity size is φ 370 × 795
effective volume55L
Overall Dimensions600 (width) x 620 (depth) x 820 (height) mm
2. Temperature range: sterilization temperature115~135℃
Moisturizing temperature45~60℃
Dissolving temperature60~115℃
3. Sensor: Temperature sensorPT100; Bottom sensor PT100; Pressure sensor 0~5V
4. Standard cooling fan: Axial flow fan1; 1 shielded fan
5. Power requirement: single-phaseAC220V ± 10%, 50Hz; Current: 17~21A
6. Pressure vessel related standards: Outer box material: Spray coated steel plate
design pressure0.241Mpa
pressure gauge range-0.1~0.4Mpa
Container material:S30408, Corrosion resistant stainless steel
national standard:GB150、YY1007
Pressure vessel category:Class I
Manufacturing License Level:D1
Weight (whole machine)90KG
Drainage pipe diameter:φ15mm
Heating power:4KW
7. Accessory: Drainage hose1 piece, 1 front cooling bucket, 1 rear cooling bucket, 1 rear cooling bucket bracket
8. Standard product: Sterilization basket-Stacked to accommodate 3; Object temperature sensor PT100
9. Safety device: safety valve-Exceeding 0.235Mpa for exhaust; Jumping thermostat - limiting the temperature at the bottom of the barrel
10. Operation mode: Normal sterilization-Can sterilize liquids such as water, culture media, reagents, and liquid food; Sterilization and insulation - items can be insulated after sterilization; Sterilization of instruments - flasks, beakers, culture dishes, and other instruments that can withstand rapid decompression are used for sterilization; Dissolve and keep warm - Dissolve the culture medium and keep warm
11. Parameter settings: Parameter settings-Maximum sterilization temperature 135℃
Maximum operating pressure-0.241Mpa
Sterilization using temperature range115℃~135℃
insulation45℃~60℃
dissolve60℃~115℃
Timed range sterilization0-5 hours and 59 minutes
dissolve0-5 hours and 59 minutes
insulation0~72 hours, 59 minutes, resolution 1 minute
12. Other functions:
ØContainer with built-in steam source-Built in 4kW heating wire, ensure sufficient water is poured in each operation
ØRapid cooling of steam emissions-Structure to avoid boiling of cooling steam with water
Øaudio prompt-There is a corresponding 'beep' sound prompt at the beginning and end of each stage of operation
Øaltitude-The manual states that it can be used at an altitude of 0-3000m, but in reality it can be used at an altitude of 0-4000m
ØSystem time setting-Change the internal time of the control system
ØSterilization counting function-After the sterilization stage is completed in the sterilization mode with the word 'sterilization', the count will be automatically accumulated
ØQuick cooling function-1 axial flow fan and 1 cover pole fan are turned on during the cooling stage to quickly cool the temperature inside the container
ØComes with a water bucket-Easy to move and avoid steam discharge
ØEasy to add water-Open the top cover and put in sterilization water
Ødischarge outlet-A drainage outlet with a diameter of 15mm facilitates the discharge of sterilization water
ØControllable steam emission temperature-Steam emission temperature regulation range: boiling point~(boiling point+45 ℃)
13. Alarm function (common alarms and their causes)
ØAbnormal temperature sensor-3 PT100 sensors malfunction
ØThe water level is too low, please replenish water-Dry burning, lack of water, air or water leakage, blocked exhaust port, abnormal bottom temperature sensor triggering alarm
ØAbnormal heating-Above 139 ℃
ØCheck the temperature sensorIs the head position correct - above 140 ℃
ØThe door lock cannot be locked-The temperature inside the container is too high, and the top cover cannot be locked tightly
ØThe door lock cannot be unlocked-The handle travel is limited and the lock cannot be tightened
ØAbnormal high pressure -The pressure exceeds 0.33Mpa, and the pressure value during the sterilization stage is higher than the corresponding value in the pressure temperature comparison table, which is 0.04Mpa
ØAbnormal low pressure -The pressure value during the sterilization stage is 0.04Mpa lower than the corresponding value in the pressure temperature comparison table
Øunheated-The temperature has not risen for 2 minutes
ØMain control board malfunction -Main control board malfunction
14. human-computer interaction
ØQuick opening mechanism-Press the top cover with one hand and swipe the handle left and right with the other hand to open and close the top cover
ØHidden front cooling bucket-Beautiful design
ØCooling bucket water level observation window-The observation window is located on the front side of the machine for easy observation of water level
Ø5-inch true color touch screen - displays temperature, pressure, and time on the same screen at a glance
Øtouch operation-Consistent with mobile phone operation, convenient to set up
15. intelligent management
ØSimple parameter setting-The interface is like a mobile app interface for inputting parameters, simple and direct
Øsmart chip-Microcomputer based intelligent process control
ØPower outage parameter memory-After saving the parameters, turning off the power can also save them without losing them
ØProgram Stage Tips-At the beginning of each stage, there is a sound prompt, and the touch screen has corresponding progress curve color differentiation
ØSleep function-Failed to lock the upper cover after timeout
Ø16 preset programs - each mode contains 4 programs, totaling 16 programs
16. hierarchical management
Øpassword permission-Prevent maintenance personnel or technicians from mistakenly changing parameter settings
Øcalibration accuracy-Calibration accuracy 0.1 ℃
17. intelligent management
ØOne click quick execution-Dealing with a large number of repetitive operations to avoid errors caused by repetitive operations
ØAppointment scheduled start-Accurate to year, month, day, and hour, convenient for adjusting working hours
18. Pharmaceutical factories and laboratories
standardG1/2 connector - Reserve 2 standard G1/2 threaded interfaces, which can be connected to temperature and pressure monitoring equipment to meet GMP and GLP requirements
19. Internet
RS485- Reserved 485 interface for remote control according to requirements
20. closed-loop control
Sleep function-Ensure the integrity of the action from placing objects to removing them: After sterilization is completed, the touch screen displays the real-time temperature inside the container, the countdown is zero, and if the lid is not opened for a short period of time, it automatically locks the lid to prevent other personnel from operating it incorrectly; Touch the real-time temperature area on the screen, unlock the upper cover, and extract the sterilized material. The real-time temperature and countdown on the touch screen will change to the last set parameters.
Panasonic high-pressure sterilization potMVS-63
IIIproductqualification
Medical Device Manufacturing Enterprise License
Medical Device Registration Certificate of the People's Republic of China
ISO9001 Quality Management System Certification
ISO14001 Environmental Management System Certification
ISO13485 Medical Device Quality Management System Certification
OHSAS18001 Occupational Health System Certification