PEC-D301 digital energy meter calibrator

NegotiableUpdate on 03/31
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
PEC-D301 digital energy meter calibrator
Product Details

1.jpg

Implement multiple modes such as virtual load verification, real load verification, and virtual verification.

Communication fault simulation testing can be conducted, such as frame loss, communication delay, channel abnormalities, etc., to comprehensively assess the communication function and fault tolerance of digital energy meters.

Can be verified during actual load verificationIEC61850-9-1/ IEC61850-9-2Analyze network packets, such as network addressesASDUNumber, sampling rate, synchronization method, etc.

Flexible configuration during virtual load verificationASDUquantity,Communication parameters such as sampling points.

Users can edit any automatic verification scheme themselves.

Having dualSCFiber optic Ethernet and dual interfacesRJ45Ethernet interface improves reliability and facilitates the integration of metering devices with different interfaces.




2.jpg

input voltage

Voltage measurement range, accuracy 0-100V, (20-100V) 0.05% RD

input current

Current measurement gear

200mA、 1A、 5A、10A

Accurate measurement range

(50mA~10A)0.05%RD

input power measurement

Accuracy of active power measurement

0.05%RD PF=1(50mA~10A)

0.1%RD PF=0.5L,0.5C(50mA~10A)

Accuracy of reactive power measurement

0.2%RD PF≥0.5

frequency measurement

Frequency measurement range and resolution

45Hz~65Hz、0.001Hz

Frequency measurement accuracy

0.01Hz

Electric energy measurement

Accuracy of active energy measurement

0.05%RD PF=1(50mA~10A)

0.1%RD PF=0.5L,0.5C (50mA~10A)

Accuracy of reactive power measurement

0.2%RD PF≥0.5