Overview

The PACON4200 online hardness analyzer adopts the titration colorimetric principle, with a compact structure, easy operation, and accurate measurement. It is an entry-level choice for water softening systems and boiler room water quality monitoring. Measurement parameters: total hardness Typical applications: boiler water supply and boiler water, circulating water, process water

Product Details

characteristic

PACON4200 online hardness analyzer Jepp instrument


●Automatic measurement and automatic cleaning

●Graphic backlit LCD display, Chinese and English menus

●External signal input control measurement, used for starting or stopping analysis externally

● 4-20mA output (optional RS485 communication module)

●2-channel relay output

● Display unit ppm CaCO3

● Continuous measurement or interval measurement (5-30 minutes)

● Optional SD card data storage (historical data, fault records)

●Less maintenance workload




specification

PACON4200 online hardness analyzer Jepp instrument

Technical parameters:

Measurement range: 0.20 - 500ppm CaCO3(See reagent type)

Measurement time: about 3 minutes, depending on the hardness of the water and the set flushing time

Analysis cycle: continuous measurement/interval measurement (5-30 minutes)/external start signal

Rinse time: 5-1800 seconds (default 120 seconds)

Water consumption: about 1-2L/analysis depends on inlet pressure and set flushing time

Display: Backlit LCD displays graphics and numerical values

Unit ppm CaCO3

Current output: 0/4-20mA, Max. 750 Ω

Relay output: 2-channel passive relay output NC, 30VDC 1A

Input: External switch signal start analysis/flow sensor

Basic Parameters:

Measurement principle: titration colorimetric method

Environmental temperature: 5-45 ℃

Water sample temperature: 5-40 ℃

Water sample pressure: 0.5-5 bar, recommended 1-2 bar

Water quality requirements: colorless, no suspended solids, no bubbles

pH 4 - 10.5, Iron:<3ppm, copper:<0.2ppm

Aluminum:<0.1ppm, manganese:<0.2ppm

Inlet/outlet connection: 1/4 'outer diameter hose

Humidity: 20-90% RH, indoor wall mounted installation

Power supply: 24VDC, 25W (optional 220VAC power adapter)

Size/Weight: 250 × 360 × 110mm, approximately 2.5Kg (including outer box shell)