- Phone
-
Address
Building 1, No. 57 Guangheli West Road, Chaoyang District, Beijing
Beijing Huayu Zhongyi Technology Co., Ltd
Building 1, No. 57 Guangheli West Road, Chaoyang District, Beijing
HYPZoil productsMulti parameter onlinetestanalyzer
I. Overview
HYPZoil productsMulti parameter onlinetestThe measurement principle of the analyzer is that when the probe undergoes high-frequency torsional vibration in the fluid,

Due to different moisture contents in oil products, the resonance frequency and amplitude of the system attenuate, and the degree of attenuation corresponds to the density, viscosity, and moisture content of the object;According to the corresponding relationship, the viscosity, density, and moisture content of the oil can be distinguished.In the formula F ∝√ ρ η, where η is the viscosity of the fluid to be measured and ρ is the density of the fluid to be measured, onlinetestThe analyzer mainly consists of two parts: sensors and instrument circuits. Sensors are bidirectional sensors that can achieve the mutual conversion of mechanical quantities and electrical signals, turning complex mechanical systems into circuit systems that are easy to analyze. At the same time, they take advantage of the easier measurement of current and voltage compared to mechanical quantities, which can have a significant impact on subsequent theoretical analysis and experiments. The sensor design structure can be equivalent to a single degree of freedom second-order system, mainly composed of sensitive components、It is composed of contact probe, spindle, transmission mechanism, electromagnetic mechanism and drive circuit, torsional vibration mechanism, crossbar, amplification sensitive element at resonance frequency, elastic torsion tube, differential coil, frequency measurement mechanism, etc. The excitation current drives the torsional vibration mechanism to resonate, driving the spindle to produce torsional vibration. This is transmitted to the sensitive element through the spindle transmission mechanism, causing the sensitive element, spindle, and crossbar to vibrate synchronously. The changes in viscosity and density are converted into changes in the shear stress torque of the sensitive element, and the amplitude and frequency of the sensitive element's motion will also change accordingly. The amplitude and frequency changes will be fed back to the crossbar through mechanical construction, amplifying the sensitive element. The microcomputer measures the amplitude and frequency of the crossbar through a differential transformer displacement sensor.
When probes such as cylinders, balls, and rings rotate freely in a fluid, their amplitude will be attenuated by the viscous torque, and the attenuation coefficient is related to the viscosity of the fluid. The fluid vibrating body is selected as the cylinderThe formula for calculating viscosity is:
Viscosity:μ=f/4πh/(1/R)12 -1/R22)Among them,

μ: Fluid viscosity
F: Fluid resistance
h: The height of the cylinder
R1: Outer diameter of cylinder
R2: Inner diameter of cylinder
densityD=k1f3+k2f2+k3f+k4
f: Resonant attenuation frequency
K1, k2, k3, k4 fitting coefficients
Moisture content:V=(D-mixed - D-oil)/(D-water - D-oil)
D oil: pure oil density
D mixture: density of water containing oil
Suitable for online monitoring of viscosity and density of lubricating oil, gasoline, diesel, hydraulic oil, gearbox oil, aviation fuel, kerosene, biodiesel, ethanol, chemical reagents, coatings and inks, and detection of material moisture content.
Food processing, laboratory analysis, fuel storage and transportation processes; Refining, petroleum industry, wind power, thermal power generation and other power generation industries; Aviation, navigation, railway and other transportation industriesMarine engineering such as drilling platforms, large-scale engineering machinery and equipment, transportation vehicles, fuel storage and transportation vehicles, oil treatment equipment, chemical experiment analysis, production process quality management, oil treatment and filtration systems; Medical equipment for coatings, inks, and printing industries
Viscosity range:0-1000 cP
accuracy:±0.5%
Repeatability:±0.3%
Density range:600-1200kg/m3
accuracy:0.1%
water content:<10%
accuracy:0.2%
Temperature measurement range:0-150℃
Temperature accuracy:0.2%
Sensor pressure range:<6.4 Mpa
(6.4 Mpa or above customized)
Sensor temperature resistance range<150℃
Signal response time:2s
Material316 stainless steel (standard);
Optional materials for processing:Standard: Highly polished with PTFE coating
Environmental level:IP68
power supply24V DC(Customer provided)
output:MRejection
Protection levelIP67
Explosion proof standard:ExdIIBT6
Calibration: Standard Sample Solution
Viscosity unit: arbitrarily set
Connection: flangeDN50,PN<4.0MPa,
threaded connectionM30×1.5
User's Choice
Flange standard:HG20592
On site viscosity sensor communication interface:RS232/RS485
Network terminal communication interface: Ethernet
Display mode: Color LCD touch screen7'
Display range:0~9999.9(Viscosity, density, moisture content, temperature)
Data access:Ethernet
Installation method: wall mounted, disk mounted
Chassis size:280mm*220*120mm
5Install

VIinterface display