E-mail
18618288790@163.com
-
Phone
18618288790
-
Address
Room 1008, Yuanli Building, No. 23 Huixin East Street, Chaoyang District, Beijing
Beijing Tian'enhantou Technology Co., Ltd
18618288790@163.com
18618288790
Room 1008, Yuanli Building, No. 23 Huixin East Street, Chaoyang District, Beijing
OMEC European and American Laser Particle Size Analyzer - DP-02 TypeIt is specifically designed for measuring the particle size distribution of fog droplets, with an open measurement interval of variable distance. The instrument can change its structure according to the application site conditions, such as placing the light source emitter and detector on an integrated guide rail or truss, or placing them separately on independent countertops or mobile devices to meet the testing needs of different application scenarios. The new version of DP-02 spray particle size analyzer can be equipped with guide rails and accessories related to auxiliary collimation devices to assist users in rapid instrument erection and optical collimation, shorten the instrument preparation time, and further ensure the reliability of test results.
Purpose:Measure the particle size distribution of liquid spray or atomized powder.
Working Principle:By utilizing the scattering phenomenon of particles on light, the instrument calculates the particle size distribution of the measured particles by measuring the distribution of scattered light energy and inverting it based on the Mie scattering theory.
OMEC European and American Laser Particle Size Analyzer - DP-02 Type
Technical Indicators:
1. Test range: 0.5-1500 μ m
2. Repetitive error:<3%
3. Sample testing cycle: 1-2 minutes
4. Measurement time: 0.5 seconds, capable of continuous sampling and measurement
5. Number of independent detection units: 48
6 Light source: He Ne laser, power 2mW, wavelength 0.6328 μ m
7. Working environment: Temperature: 5-35 ℃, humidity:<85%, Fourier lens has no direct light interference
Output project
Particle size distribution table, particle size distribution curve, average particle size, median size and other characteristic particle sizes, specific surface area, etc
Product Features:
1. Specially designed for the parallel light path of spray particle size measurement, the instrument structure can be changed according to the application site conditions.
2. Optional auxiliary collimation device accessories can quickly achieve instrument alignment, combined with an automatic centering mechanism to complete instrument preparation work in a very short time.
3. The light source adopts a 2mW 0.6328um He Ne gas laser emitter, which has the advantages of good monochromaticity, high coherence, small divergence angle, and strong stability.
4. Adopting an integrated laser emitter design effectively reduces the impact of thermal deformation of the laser tube and external mechanical vibration on the stability of the instrument.
5. It includes manual fine-tuning and intelligent automatic centering functions, achieving fast and accurate optical centering and ensuring measurement repeatability.
6. Universal 82mm UV lens protection device is optional to avoid possible pollution of spray to the main optical components of the instrument, which is easy to clean and replace.
7. Randomly attached disc standard samples can be used to calibrate and verify the instrument at any time.
8. Measurement time and frequency can be set, multiple measurement results can be continuously recorded, and it has statistical comparison and analysis functions.
Software functions:
1.2 Distribution Models: Universal Distribution, Single Peak Distribution
2. Test report mode: Various report templates such as regular reports, statistical reports, trend reports, etc., which can be customized.
3. Cumulative distribution direction of granularity: from small to large or from large to small
4. Particle size distribution results: Volume distribution, surface area distribution, and quantity distribution can be converted
5. Analysis software language: Chinese, English
6. Particle size range can be customized
7. The test report can be exported in Excel, Word, PDF or other text formats, and all images can be exported by right clicking the mouse
8. Multiple test reports can be opened simultaneously for comparison of data between different test reports
9. The test report items can be input and modified according to industry requirements, such as characteristic particle size, specific particle size values, etc., and can be set in a fixed report format.
Attachment: DP-02 Test Report (see next page)