- Phone
-
Address
New City Science and Technology Industrial Park, Xincheng District, Xi'an City
Shaanxi Fusaisi Instrument Co., Ltd
New City Science and Technology Industrial Park, Xincheng District, Xi'an City
Big DipperNon contact online material moisture meter T-BD5-MS1204DpPowder/Particle Online Moisture Testing Instrument
1.MS1510BOnline moisture testing transmitter
Install the MS1510B moisture sensor antenna and connector directly at 10cm above and below the conveyor belt with a constant flow rate. Install the transmitter at a suitable location within 6 meters of the conveyor belt. The transmitter outputs a standard signal of 0-20mA and is connected to the L-BD3m moisture measurement and control device through a cable. The measuring and controlling device can be installed in the instrument room or other indoor locations.
After calibration, the instrument can be tested. Online calibration devices should be installed in important occasions.
MS1510B horn type: accuracy:<± 0.5% (affected by flow changes); Measurement range: 0-60%
A conveyor belt with significant flow fluctuations must be equipped with a BD6201-X-ray densitometer for end face density correction. If the user already has a nuclear scale or electronic belt scale, their instantaneous flow signal can also be used as correction. The standard requirement for flow signal is 4-20 or 0-20mA. Accuracy ± 1.5%
Essential components:
Moisture transmitter MS1510B V3 Microwave attenuation detection technology includesBD5CMD.On site display,calibration
Technical * Components:
Flat feeder |
A mechanical baffle that can level the surface of the material, facilitating testing and protecting the probe |
X-ray densitometer |
6201 densimeter+Flow velocity detection. Perform end face density correction |
Radiation source protection installation cover |
When installing a radiation flowmeter, protective measures must be taken |
humidity correction |
Correct the impact of moisture content in the atmosphere or production air *bigImpact 1.5~2% |
2. Application of MS1204Dp conveyor belt/sand mixer moisture testing
1)DxSeries Product Series
T-BD5-MS1204Dp(Vertical) transmitter
Suitable for mixing machinesConveyor belts, etc Intercept sampling
Technical specifications:
Sample speed:<3m/s
sample thickness:>30mm
Sensing surface size:φ72mm
range: 0- 30% typical, up to 60%
resolution: 0.01%
repeatability accuracy: 0.01 or 1% of reading,which ever is great.
accuracy:<1% FS
response time:<100ms
2)T-BD5-SMS1204Dh(Integrated) Flat bottomed transmitter
Suitable for kettle wallsInstallation of sliding chute
Technical specifications:
sensing surface:φ85mm
Standard Blue Size: 150mm
Sample speed:<10m/s
Sample particle size:<10mm
sample thickness:>10mm
range: 0- 30% typical, up to 60%
resolution: 0.1%typical, best to 0.01%
repeatability accuracy: 0.01 or 1% of reading,which ever is great.
accuracy:<1% FS
response time:<100ms
3)A013BD5-SMS1204Dh(Separate) Flat bottomed transmitter
For separate installation
can,container,Sliding chute,hopper,stove,The performance of dryers and other applications is the sameT001-BD5-MS1204Dh
DpType usage and installation diagram

Introduction to transmitter functionsRefer to "Introduction to BD4&5 Intelligent Transmitter Measurement and Control Device"
Essential components:
equipment |
model |
range |
configuration |
Moisture transmitter |
MS1024Dp-0.5 |
0.5->30% |
containBD5CMD.On site display,Operable calibration |
Moisture transmitter |
MS1024Dp-0.1 |
0.1->30% |
containBD5CMD.On site display,Operable calibration |
Moisture transmitter |
MS1024Dp-0.01 |
0.05->10% |
containBD5CMD.On site display,Operable calibration |
3. MS1204CCUse of clamp type conveyor belt
CCType sensor is composed of2Composition of block plates,Suitable for online testing of conveyor belts such as paper, wooden boards, plastic boards, plywood, tape, etc., for fixed thickness boards.
Uniform thickness of the flat plate,If the movement is stable, the testing effect is good. Material thickness at the probe location10The material state must be stable within a distance of several times. There must be no other
machineElectrical equipmentSet.
The error caused by thickness or distribution flatness is approximately equal to:[Relative change rate of thickness%/50 ](%)
Different types of materials,Different calibration tables must be used.
Technical specifications
General repeatability accuracy: ±0.2 to±1.0 %
Moisture testing range:0-30%
Operating temperature:-20-100℃
ambient humidity:0-90%RH
width: 0.20-1.7m(Customized according to user needs),Width exceeds1.2rice,each10Centimeter increase100Yuan)

Essential components:
Moisture transmitter MS1024cc containBD5CMD.On site display,calibration
Humidity correction:Correcting the impact of moisture content in the atmosphere or production air. *Big impact1.5~2%
Big DipperNon contact online material moisture meter T-BD5-MS1204DpPowder/Particle Online Moisture Testing Instrument
4.MS4300IR100/200 Infrared moisture tester
Fast and online analysis
„ System features Characteristics
zNo need to prepare samples, non-contact testing, non-destructive testing, and will not introduce errors in other volatile components;
zOnline real-time testingUsed for online continuous testing and process control
zSampling calibration using regression analysis is suitable for your system
zUp to100 calibration channels, suitable for various samples
zLCD displays test results
zAutomatic temperature and humidity compensation
„ system performance Features
zThe range of moisture testing is0-90%
zprecision±0.1%, actual accuracy±0.5%, maximum error<±1%.
zSampling signal processing adopts program amplification circuit
zAlarm settings and switch control outputs
zstandardRS232C and 4-20mA signal output
zAutomatic system recording 24 Hourly on-time average test value,Equipped with a printer to print various data.
„ Technical performance of measurement and control device PROCESSOR SPECIFICATIONS
input signal |
0-5Vor4-20mA |
output signal |
0-5Vor4-20mA |
Test Principle |
Near infrared reflection method |
LCDdisplay |
moisture content;channel number;time |
Calibration method |
nMoisture setting affects the sampling signal value; |
oMoisture setting for sampling calculation values |
|
Linear processing |
2 or 3 Sub fitting curve,Or line insertion calculation |
Output measurement |
arbitrary,Adjustable |
response time |
0.1s |
ambient temperature |
-10-70℃;10-85%RH |
data recording |
24 On time average record |
self-diagnosis |
non-reflective,Calibration error,Probe malfunction,Operation error,Motor malfunction,Light source damaged,Probe temperature exceeds the limit |
通信口 |
RS232/RS485 |
power supply |
220/110V AC 50/60Hz < 50W |
probe size |
33×17.5×20(cm) |
5. sMS2100Solid Particle Powder Transmitter Series
application
Solid Material Chemical Engineering
Humistatus/HVAC (Heating, Ventilation, and Air Conditioning) application for hot air/ventilation/air conditioning
Industrial furnace protection gas detection, chemical gas curtain moisture detection
Industrial controls for online detection of moisture in various solid materials
Indirect moisture testing of solid/liquid materials
T001-BD5CMD+sMS2123NPlug in water activity transmitterOperating Temperature: -40 to 85°C
work pressure:<0.3Mpa
insertion depth: 40- 200mmadjustable
Install Connection: ZG1” /1”NPT
2x16 LCDLCD (Liquid Crystal Display), 5x4Keyboard operation
Standard signal:0/4 to 20mA output
Communication interface:RS232/485communication, STIMcom or Modbus supportGeneral computing function configuration and support for special function modules.
full-featuredSTIMSmart transmitter.
T001-BD5CMD+sMS2123pPlug in water activity transmitterOperating Temperature: -40 to 85/100°C
work pressure:<1.0 Mpa
insertion depth: 200mm
Install Connection: ZG1” /1”NPT
2x16 LCDLCD (Liquid Crystal Display), 5x4Keyboard operation
Standard signal:0/4 to 20mA output
Communication interface:RS232/485communication, STIMcom or Modbus supportGeneral computing function configuration and support for special function modules.
full-featuredSTIMSmart transmitter.
T001-BD5CMD+MS2110pHT200Plug in water activity transmitter
Operating Temperature: -40 to 180°C
work pressure:<1.0 Mpa
insertion depth: 200mm
Install Connection: ZG1” /1”NPT
2x16 LCDLCD (Liquid Crystal Display), 5x4Keyboard operation
Standard signal:0/4 to 20mA output
Communication interface:RS232/485communication, STIMcom or Modbus support
General computing function configuration and support for special function modules.
full-featuredSTIMSmart transmitter.
T001-BD5CMD+MS2110pHT300Plug in water activity transmitter
Operating Temperature: -40 to 300°C
work pressure:<1.0 Mpa
insertion depth: 200mm
Install Connection: ZG1” /1”NPT
2x16 LCDLCD (Liquid Crystal Display), 5x4Keyboard operation
Standard signal:0/4 to 20mA output
Communication interface:RS232/485communication, STIMcom or Modbus support
General computing function configuration and support for special function modules.
full-featuredSTIMSmart transmitter.