Nitrogen dioxide NO2 detection tube Indoor gas detector Poison detection tube

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Nitrogen dioxide NO2 detection tube Indoor environmental gas detector Poison detection tube NO2 detection tube NO2 detection tube is a sealed slender glass tube that presents a specific color due to being filled with specific reagents and coloring agents. When a gas sample containing nitrogen dioxide enters the detection tube, a chemical change occurs, causing a change in color. The scale corresponding to the boundary of observing color changes is the concentration of nitrogen dioxide in the tested sample.

Product Details

Nitrogen dioxide NO2 detection tube Indoor environmental gas detector Poison detection tube

Nitrogen dioxide detection tube NO2

The nitrogen dioxide detection tube is a sealed, slender glass tube that presents a specific color due to being filled with specific reagents and coloring agents. When a gas sample containing nitrogen dioxide enters the detection tube, a chemical change occurs, causing a change in color. The scale corresponding to the boundary of observing color changes is the concentration of nitrogen dioxide in the tested sample.


9L specification:

Testing Scope 0.5-30ppm 30-125ppm
Extraction frequency 2 times (standard) (200ml) Once (100ml)
correction factor 1 According to the correction table
pumping time 1 minute 30 seconds

Detection line: 0.1ppm (pumping twice)

Color Change: White, Yellow, Orange

Temperature and humidity correction: No temperature correction required

Relative standard deviation: 10% (0.5-3ppm), 5% (3-30ppm)

Validity period: 36 months

reaction principle:NO2+o-Ortho toluidineyellowreactant

This detection tube can detect other gases

gas name conversion factor Extraction frequency Testing Scope
iodine 0.4 2 0.2-12ppm



9P specification:Suitable for indoor air testing (at 20 degrees Celsius, 0.24mg/m3 is approximately equivalent to 0.125ppm)

Testing Scope 0.02-0.2ppm
Flow setting 100ml/min(3000ml)
conversion factor 1
Detection Time 30min

Detection line: 0.004ppm (3000ml)

Color Change: White, Yellow, Orange

Temperature and humidity correction: No temperature correction required

Relative standard deviation: 10% (0.02-0.06ppm), 5% (0.06-0.2ppm)

Validity period: 30 months

reaction principle:NO2+3,3 ', 5', 5-Tetramethylbenzidine Orange Yellow Reaction Product

DSC_2407



Nitrogen dioxide NO2 detection tube Indoor environmental gas detector Poison detection tube