- Phone
-
Address
Room 608, Building 1, No. 348 Huajiang Road, Jiading District, Shanghai
Shanghai Fatai Precision Instrument Co., Ltd
Room 608, Building 1, No. 348 Huajiang Road, Jiading District, Shanghai
| Type B | |
| Sensor type: | Aluminum oxide capacitive sensor |
| SADPminiEx Handheld Dew Point MeterIntrinsic safety: |
|
| Dew point range: | Silver dot: Dew point from -110 º C to -20 º C (-166 º F to -4 º F) |
| Purple dot: Dew point from -100 º C to 0 º C (-148 º F to+32 º F) | |
| Red dot: Dew point from -80 º C to -20 º C (-112 º F to -4 º F) | |
| Grey point: Dew point from -80 º C to 0 º C (-112 º F to+32 º F) | |
| Blue dot: Dew point from -80 º C to+20 º C (-112 º F to+68 º F) | |
| Display unit: | Celsius dew point |
| º F - Fahrenheit Dew Point | |
| Ppm (v) - parts per million (volume ratio) | |
| Ppb (v) - One billionth by volume | |
| Ppm (w) - parts per million (mass ratio) | |
| g/m3-Grams per cubic meter | |
| Lbs/MMSCF pounds per standard million cubic feet | |
| Display: | Green background with blue text, 128 x 64 pixels, graphic LCD with LED backlight |
| Sensor calibration accuracy: | ± 2 ° C dew point |
| Automatic calibration: | Electronic 'range check', simple menu operation, password protected to prevent unauthorized operation |
| Repeatability: | Better than ± 0.2 ° C dew point |
| power supply | 3.6V, 2.3Ah rechargeable lithium battery, including charger |
| Battery life: | Can work continuously for more than 80 hours after being fully charged at 20 º C/68 º F |
| *Full of time: | 14 hours |
| Pressure correction: | The built-in calculator displays the dew point pressure, and the gauge pressure can be displayed in kPa, kg/cm2, bar, or psi |
| temperature coefficient | Temperature compensation for working range |
| Sample gas connection: | One end is a 6mm or 1/4'Swagelok stainless steel card sleeve connector, and the other end is a 6mm ID hose stainless steel quick nozzle, |
| Randomly provide 1 2-meter-long 6mm ID PTFE tube | |
| Electrical connection: | 9-pin D-type, used for 4-20mA analog output, RS485 serial communication, and PC interface, charger with independent socket |
| Operating Temperature | -10 ° C to+50 ° C |
| Storage temperature and humidity: | -40 ° C to+80 ° C/95% RH, non condensing |
| working pressure: | 大30 kPag (0.3 barg / 4 psig) |
| Working humidity (external): | 95%RH, non-condensing |
| Typical response time: | From dry to wet (-10 ° C to -60 ° C): less than 120 seconds |
| From wet to dry (-110 ° C to -20 ° C): less than 20 seconds | |
| Sample gas flow rate: | Recommended 2 to 5 l/min, with a maximum of 10 l/min |
| Replace desiccant: | Can be replaced on-site |
| Sensor lifespan: | 5-10 years (depending on application) |
| Replace sensor: | Can be replaced on-site |
| safety | Multi level password protection |
| Preheating time: | 10 seconds |
| Protection level | IP54 / NEMA 12 |
| Includes attachments: | 2-meter PTFE sampling tube, universal battery charger, user manual |
| F型 | Type B+the following functions: |
| Analog output: | External power supply 4-20mA circuit |
| Includes attachments: | 4-20mA connector for analog output |
| L-shaped | Type B+the following functions: |
| Data recording: | 16000 sample data with date and time, stored in selected measurement units, available for download on PC |
| Data unit: | 20 independent units |
| Sampling rate: | From once every 6 seconds to once a day, users can choose |
| The data shows that: | Digital and graphic display |
| Real time recording: | Directly connect to PC to monitor, record and graphically display data |
| Serial communication: | RS485, Baud rate 9600- half duplex |
| Analog output: | External power supply 4-20mA circuit |
| Includes attachments: | 4-20mA/RS485 connector for analog output and serial communication; Used for "real-time data recording" and download count |
| According to the serial port or USB interface (including cable) of the PC; Software CD | |
| System requirements: | Serial port connection - Windows 95 and above; USB connection - Windows 2000 and above |