Nanjing Temperature and Humidity Test ChamberTechnical parameters and performance indicators:
High and low temperature damp heat test chamber |
GDS-50 |
GDS-100 |
GDS-150 |
GDS-225 |
|||||
GDS-408 |
GDS-800 |
GDS-010 |
GDS-015 |
||||||
Studio size (mm) |
300×400×450 |
400×450×550 |
500×600×500 |
500×600×750 |
|||||
600×800×850 |
800×1000×1000 |
1000×1000×1000 |
1000×1500×1000 |
||||||
Operation mode |
Fixed value operation, program operation |
||||||||
|
skill technique ginseng number |
temperature upper limit |
+100℃/+150℃/+200℃ |
|||||||
lower temperature limit |
0℃、-10℃、-20℃、-30℃、-40℃、-50℃、-60℃、-70℃、-80℃ |
||||||||
temperature fluctuation |
±0.5℃ |
||||||||
Temperature Uniformity |
≤2.0℃ |
||||||||
temperature deviation |
±2℃ |
||||||||
humidity range |
20~98%RH |
||||||||
humidity deviation |
+2/-3%RH |
||||||||
warm-up time |
40min(20℃~100℃) |
||||||||
Cooling time |
25min(20℃~0℃)、40min(20℃~-20℃)60min(20℃~-40℃)、80min(20℃~-70℃) |
||||||||
Temperature and humidity controller |
Imported temperature and humidity controller |
||||||||
temperature sensor |
Pt100 platinum resistor |
||||||||
humidity sensor |
Pt100 platinum resistance dry wet bulb |
||||||||
|
manufacture cold be unify |
refrigeration unit |
Imported fully enclosed French Taikang compressor unit (environmentally friendly) |
|||||||
Cooling method |
Single stage compression refrigeration, binary cascade compression refrigeration |
||||||||
cooling method |
Air cooling/water cooling |
||||||||
heater |
Nickel chromium alloy electric heater |
||||||||
Humidification method |
Boiler type evaporation |
||||||||
humidifier |
Stainless steel heating tube |
||||||||
Dehumidification method |
Low temperature dehumidification |
||||||||
circulating fan |
axial flow fan |
||||||||
Security protection measures |
Overtemperature, overload, overpressure, oil pressure Protection against leakage, water interruption, short circuit, power phase loss, and phase error |
||||||||
|
material material |
housing material |
Cold rolled steel plate electrostatic spraying |
|||||||
Inner wall material |
SUS304 stainless steel plate |
||||||||
thermal insulation material |
Ultra fine glass wool |
||||||||
Standard Configuration |
One test hole of size 50, two sets of test sample shelves, lighting fixtures, and insulated glass observation windows coated with electric heating film |
||||||||
Optional accessories |
Communication interface, remote monitoring computer and software |
||||||||
power supply |
380V±10%,50Hz,3P+N+G |
||||||||
product standard |
GB10586: Technical Conditions for Damp Heat Test Chamber GB10589: Technical Conditions for Low Temperature Test Chamber |
||||||||
Test Method |
GB/T2423.1、GB/T2423.3、GB/T2423.9、GJB150.4 |
||||||||
|
|||||||||
Nanjing Temperature and Humidity Test ChamberTechnical parameters and performance indicators:
precision range : Temperature±0.1℃Indication accuracy: temperature±0.1℃Resolution:±0.1℃
temperature sensor:Platinum resistancePT100Ω/MV
circulatory system:Temperature resistant and low-noise motor, single cycle, extended Shaft, stainless steel multi blade centrifugal fan blade
Heating method:Nickel chromium alloy electric heating heater
Cooling method:Unit cooling method/dual element(repetition)Cooling method (air cooling)
cooler:France“Taikang”Multi stage membrane evaporator
compressor:France“Taikang”Fully enclosed refrigeration compressor/France“Taikang”Fully enclosed refrigeration compressor unit
timer function: 0.1~99.99(STheMTheH)
Security device:Power overload, leakage, grounding, over temperature protection (with voice prompt), fast fuse, compressor overvoltage protection
Standard Configuration:High and low temperature resistant tempered glass internal heat insulated glass observation window:220×180(mm)one,One lighting lamp
Test lead hole ¢50mmOne, sample rack2Pay, Equipped with casters
Equipment structure
The equipment mainly consists of a box, a heating system, a cooling system, an air circulation system, and a control system.
The design of the test box is made by CNC machining and adopts non reactive handles, making it easy to operate.
The sealing of the box door is made of refined silicone rubber, so that there is no aging or hardening phenomenon at high and low temperatures.
The inner liner is importedSUS304Stainless steel plate, with a shell ofA3The board is treated with spray coating, making it smoother and more aesthetically pleasing.
The insulation system uses ultra-fine glass fiber or hard polyurethane foam to fill the insulation area to ensure the temperature inside the box.
The independent heating and cooling system enables the equipment to heat up and cool down more effectively, and the cooling system is a fully automatic control and safety protection coordination system.
Adopting a multi wing strong air circulation system to avoid any dead corners and ensure even temperature distribution within the testing area.
The air circulation design ensures that the air pressure and speed meet the testing standards, and can accelerate the temperature recovery time at the moment of opening the door.
*South Korea'TEMI880'Programmable temperature and humidity controller(5.7inch16True color touch screen)Chinese style menu operation method, memory program group120Group, each program group contains program segments1200Segments, all loops, the most abundant999return
The equipment meets the standards
GB/T2423.1-2001Basic Test Procedures for Electrical and Electronic Products experimentA:Low temperature test method
GB/T2423.2-2001Basic Test Procedures for Electrical and Electronic Products experimentB:High temperature test method
GB/T2423.3-2006Environmental testing methods for electrical and electronic products experimentCabConstant Damp Heat Test
GB2424.1-89Basic Environmental Testing Regulations for Electrical and Electronic Products Guidelines for High Temperature and Low Temperature Testing