E-mail
ceas@naide.cn
- Phone
- Address
Chongqing Naide Automation Technology Co., Ltd
ceas@naide.cn
NIPm-5000-DR Water Quality Online Automatic Monitoring and Quality Control Instrument

execution standard
HJ915-2017 Technical Specification for Automatic Monitoring of Surface Water Quality (Trial)
HJT 372-2007 "Technical Requirements and Testing Methods for Water Quality Automatic Samplers" HJ 354-
2019 Technical Specification for Acceptance of Online Monitoring System for Water Pollution Sources (CODCr, NH3-N, etc.)
(HJ 354-2019 replaces HJ_T 354-2007)
working principle
Through different quality control methods (immediate quality control, periodic quality control Timed quality control) and quality control modes (standard sample verification, parallel sample testing, spiked recovery rate testing, standard one standard sample) Verification, standard sample verification, blank standard sample verification, dynamic standard sample verification, dynamic spiked recovery) can provide different concentrations of standard samples for online monitoring instruments to obtain their online monitoring data, through wired or wireless means Transmit it to remote control software in a wired mannerIt can remotely check whether the online water quality monitoring instrument is working properly, whether the data is biased, and whether it is effective, achieving remote automatic monitoring of the accuracy of the online water quality monitoring instrument data.
Product Features
Distributed layout, with quality control terminals placed next to online monitoring instruments, eliminating the need for a whole space;
The device is compact and easy to install;
The standard sample has a short transmission distance, minimizing transmission errors to the greatest extent possible;
〉A display controller can control 16 quality control terminals, and adding monitoring parameters in the future only requires the corresponding number of quality control terminals, making the upgrade very convenient;
Temperature controlled room for refrigerated storage of standard solution, refrigerated temperature4℃±2℃;
Electronic access control, password permission login.
It has functions such as standard sample verification, parallel sample testing, spiked recovery rate testing, blank standard sample verification, dynamic standard sample verification, and dynamic spiked recovery rate.
Scope of application
Online water quality monitoring equipment applied to sewage treatment plants, waterworks, surface water, drinking water sources, aquaculture, industry and other fields
Be prepared and used in conjunction.
working method

Main technical parameters