Welcome Customer !

Membership

Help

Chongqing Naide Automation Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Chongqing Naide Automation Technology Co., Ltd

  • E-mail

    ceas@naide.cn

  • Phone

  • Address

Contact Now

NIPm-3000COD/Ammonia Nitrogen/Total Phosphorus/Total Nitrogen Water Quality Online Analyzer

NegotiableUpdate on 11/28
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

NIPm-3000COD/Ammonia Nitrogen/Total Phosphorus/Total Nitrogen Water Quality Online Analyzer NIPm-3000 Four Parameter Water Quality Online Analyzer uses sequential injection spectrophotometry technology to sequentially add water samples, reagents, etc. into the digestion tank for fully automatic process control such as mixing, digestion, colorimetry, and cleaning. The water sample concentration is automatically calculated by absorbance before and after the reaction, achieving multi parameter quantitative analysis of COD, ammonia nitrogen, total phosphorus, and total nitrogen

Product Details

NIPm-3000COD/ammonia nitrogen/Total Phosphorus/Total nitrogen water quality online analyzer

NIPm-3000typefourParameter water quality online analyzer adopts sequential injection- spectrophotometric technology, sequentially adding water samples, reagents, etc. into the digestion tank for fully automatic process control such as mixing, digestion, colorimetry, and cleaning. The water sample concentration is automatically calculated based on the absorbance before and after the reaction, achievingCODMulti parameter quantitative analysis of ammonia nitrogen, total phosphorus, and total nitrogen.

Application fields:

·Sewage treatment plant, water supply plant

·surface water

·Aquaculture, industrial wastewater, etc

Features:

·Can measure simultaneouslyCOD、 Four parameters of ammonia nitrogen, total phosphorus, and total nitrogen, with low one-time investment

·Small footprint (approximately)0.2㎡)

·Low reagent consumption and low operation and maintenance costs

·Platform based and modular design, easy to maintain and quickly repair

Function:

·Visual interface, prompt menu, easy to operate

·Transportation operation and maintenance, easy to operate, with small maintenance volume

·Intelligent diagnosis of pipeline blockage, leakage, reagent shortage, abnormal information, and instrument failure automatically uploads alarms

·Remote quality control, remote sample verification, remote monitoring of instrument status, remote online software upgrade

·Automatic range switching, automatic verification of standard samples, automatic determination of spiked recovery rate, and automatic temperature compensation

·Zero point and range automatic calibration, waste liquid separation, and automatic cleaning

·Comprehensive user permission management and intuitive measurement


technical indicatorsTechnical index

Measurement factor

COD

ammonia nitrogen

Total Phosphorus

Total Nitrogen

Measuring principle

Potassium dichromate oxidation spectrophotometric method

Salicylic acid spectrophotometric method

Ammonium molybdate persulfate oxidation spectrophotometric method

Alkaline potassium persulfate spectrophotometric method

measurement range

(0~1000)mg/L

(0~100)mg/L

(0~50)mg/L

(0~50)mg/L

Range division

(0~1000)mg/L

(0~100)mg/L

(0~50)mg/L

(0~50)mg/L

measurement accuracy

±1%

±1%

±1%

±2%

zero drift

±3mg/L

≤ 0.02 mg/L

±3%

±2%

Span Drift

±3%

≤1%

±3%

±3%

Actual water sample comparison experiment

<±8%

<±8%

<±8%

<±8%

Low detection limit

10mg/L

0.02mg/L

0.02mg/L

0.05mg/L

Short measurement cycle

30min

20min

30min

40min

digestion time

≥500s

≥200s

≥500s

≥1200s

Quantity of waste liquid each time

6mL

5mL

5mL

6mL

Reagent replenishment interval

a month

sampling period

Time interval measurement mode (adjustable from 1 to 9999 minutes) and timed hourly measurement mode.

calibration cycle

It can automatically set and start calibration, and can be adjusted at any time.

maintenance cycle

Usually once a month, each factor takes about 5 minutes and can be remotely maintained.

output

Analog output: (4-20) mA

Digital Communication: RS232/RS485 MODBUS-RTU Communication Protocol

ambient temperature

(5-40)℃

ambient humidity

(5-80)%RH

Overall dimensions

Height x Width x Depth=1200mm x 600mm x 350mm

Protection level

IP65

weight

49kg

power supply

AC(220±10%)V,50Hz,5A

Other functions

Abnormal alarm, power outage without losing data; Automatic alarm when no water sample or reagent is collected; Touch screen display and command input; Abnormal reset, after power failure, the instrument automatically discharges residual liquid and returns to working state.

Ordering Notice:

When placing an order, please specify the on-site conditions: power supply, output, medium composition, measurement range, water sample temperature and pressure, cable length, installation method, etc.