Multi functional chemical flooding physical simulation system

NegotiableUpdate on 01/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The multifunctional chemical drive physical simulation system is easy to install and move, with high temperature accuracy and small temperature fluctuations. The system is equipped with over temperature, over pressure, power-off, and power-off protection devices, which are safe and reliable.

Product Details

Multi functional chemical flooding physical simulation systemThe device is mainly used for polymer flooding, surfactant flooding, and composite flooding research. It can conduct physical simulation experiments of reservoir displacement under simulated reservoir temperature and pressure, and obtain the effect of improving oil recovery under different types of displacement conditions. The experimental results are used to evaluate the oil displacement effect of chemical agents and optimize the injection parameters of chemical flooding schemes.

The injection pressure tracking system maintains a constant pressure difference, and the injection control mode of the device is divided into various working modes such as constant speed and constant pressure. The constant temperature box adopts a hot air stirring and segmented structure. Easy to install and move, ensuring high temperature accuracy and minimal temperature fluctuations. The system is equipped with over temperature, over pressure, power-off, and power-off protection devices, which are safe and reliable.

Multi functional chemical flooding physical simulation systemMain performance indicators:

Work pressure: 50MPa; Accuracy: 0.25 level

Working temperature: 160 ℃; Accuracy ± 1 ℃

Pump speed range: 0.001-60ml/min; Accuracy ± 0.3%

Natural rock core holder: φ 25 × 25-150mm

Measurement method: Micro pump method for automatic phase separation

Oil water measurement accuracy: ± 0.1ml

Optional image based automatic phase separation oil-water metering accuracy: up to ± 0.05ml

Power: 380V, 6KW