Technical specifications
1. Light source: DC12V22w imported lamp, equipped with power energy-saving management
2. Optical path: 8-channel vertical optical path system
3. Wavelength range: 400-1000nm (optional 340-1100nm)
4. Filter: Standard configuration includes 405, 450, 492, 630nm, other wavelengths are optional, and the filter wheel can simultaneously load 15 filters
5. Reading range: 0-5.000Abs
6. Linear range: 0-3.500Abs
7. Resolution: 0.001Abs
8. Accuracy: ≤± 0.01Abs
9. Stability: ≤ ± 0.002Abs
10. Repeatability: CV ≤ 0.2%
11. Incubation function: Room temperature+4 ℃~50 ℃, accuracy at 37 ℃, soil temperature at 0.5 ℃ (optional)
12. Vibration board function: The amplitude of the vibration board can be adjusted in 3 levels, and the vibration board time can be set from 0-255 seconds
13. Display operation: 8 o'clock color LCD screen, visual board layout, display of whole board information, touch screen operation; 14. Workstation software is randomly equipped with a professional computer terminal for online operationMicroplate ReaderAnalysis software with functions such as project detection, data processing, curve fitting, threshold judgment, quality control evaluation, database management, and custom report format
15. Interface: COM+USB port (communication, printing, USB drive)
16. Power supply: AC100-240V50/60Hz, wide voltage design, suitable for global network power supply
17. Net weight: approximately 11Kg
18. Size: Approximately 444mm (L) * 320mm (W) * 318mm (H)