Product Name:Magnetic glassy carbon electrode
Product model: CGC110
Product features: Integrated design of magnet and electrode achieved
Magnetic glassy carbon electrodeProduct List: 1 standard electrode
2. One high-strength magnet
3. Five pieces of fluororubber sealing rings
If further treatment of glassy carbon electrodes is required using ultrasound, the following points should be noted:
1. It is not advisable to place the electrode directly at the bottom of the cleaner to avoid damage to the electrode caused by excessive ultrasonic waves;
2. The ultrasound time should not be too long, generally not exceeding 30 seconds each time;
3. The container and solution used for ultrasound should be kept clean;
4. The electrodes of the PTFE jacket should avoid using high-temperature cleaning solutions for cleaning;
5. The use of strong sulfuric acid cleaning solution should be avoided for polyetheretherketone electrodes (refer to instructions: corrosion resistance of PEEK material);
The applicable range of glassy carbon electrode potential is limited, and attention should be paid to avoiding the use of excessive current or voltage;
Preservation of glassy carbon electrodes. Remove as much residue as possible from the surface of the electrode to keep it clean, and then store it in a dry and dark environment at room temperature.