Welcome Customer !

Membership

Help

Anhui Maoshu Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

MS7974 thermocouple or thermistor input signal isolator (one in one out)

NegotiableUpdate on 10/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

MS7974 Thermocouple or Thermistor Input Signal Isolator -1. Characteristics -1 in 1 out · Suitable for thermocouple or thermistor input · Input type and range can be programmed through USB configuration interface · Isolated output 4-20mA or 1-5V. 2. Technical specifications · Suitable for thermocouple or thermistor signal input, after isolation, converted into standard DC signal output to control system or other instruments

Product Details

MS7974 thermocouple or thermistor input signal isolator

1、 Characteristics

1 in, 1 out

·Suitable for thermocouple or thermistor input

·The input type and range can be programmed through the USB configuration interface

·Isolated output 4-20mA or 1-5V

2、 Technical specifications

·Suitable for thermocouple or thermistor signal input, after isolation, converted into standard DC signal output to control system or other instruments.

is an intelligent temperature transmission isolation safety barrier with a USB interface, which can be programmed through a PC and control software to set the input type and range.

·Single channel, one input, one output. Powered by an independent DC power supply, with isolation between the power supply input output.

It can be used in conjunction with unit combination instruments, DCS, PLC and other systems, and is widely used in various temperature detection and control systems in industrial production processes.

Standard 35mm DIN rail card installation. Plug and unplug terminal block with spring compression connection.


● Input

Input signal type: Thermal resistance P1100, Cu50, Cu100, etc

Thermocouples K, E, S, B, R, T, N, J, etc

Measurement range: depends on the type of sensor used

(The input signal type and range can be programmed, and the instrument has been set before leaving the factory or according to the user's requirements. Please refer to the data printed on the instrument label or casing for details.)


● Output

Output signal: 4-20mA, 1-5V, or other specified signal

Load resistance: ≤ 550 Ω (factory default) when outputting 4-20mA current

When the output voltage is 1-5V, ≥ 100K Ω

Output ripple:<10mV p-p

● Power supply

Power supply voltage range: 20~30VDC

Rated supply voltage: 24VDC

Current consumption (when powered by 24VDC):<30mA (1 in 1 out)


● Comprehensive parameters

Standard accuracy: ± 0.1% (see input type/range and measurement error table)

Wire resistance: Three wire input of thermal resistance, ≤ 10Q/wire

Influence of wire resistance: Three wire input of thermal resistance,<0.005%/Ω

Cold end compensation accuracy: ± 1 ℃ (compensation range -15~+75 ℃)

Temperature drift: ± 0.05% FS/10 ℃

Response time:<10ms

Power on stability time:<1s

Isolation capability: 2.5KV.1min.50Hz between input output power supply

Insulation resistance: ≥ 100M Ω/500VDC between input output power supply

Working environment temperature: -20-+60 ℃

Storage environment temperature: -40~+80 ℃

Shell material/protection level: PC (polycarbonate) material/IP20

Dimensions: 16 × 116 × 110 (mm)


3、 Model selection

MS7974选型表.pngMS7055选型图.pngMS7055选型图.png

4、 Dimensions and installation diagram
● Installation
1.35mm standard DIN rail card installation. Please pay attention to the stability and firmness of the card position during installation.
2. Install vertically as much as possible to facilitate heat dissipation inside the instrument.
● Typical applications
1. Used to provide distribution voltage to on-site transmitters and isolate and convert input signals.
2. Devices connected to the site: Active electrical signals from two-wire or three wire non smart transmitters.
3. Connect to the control system (or other unit combination instruments): The AI card of each DCS/PLC receives current or voltage signals.

● External dimensions
Width (thickness) x height x depth=16 x 116 x 110 (mm)

外形尺寸图.png

5、 Terminal wiring diagram
Terminal wiring: Plug and unplug terminal blocks, connected by self boosting spring type compression. The cross-sectional area of the wire shall not exceed 2.5mm2Multiple bundles or single stranded cables.

MS7974接线图.png


browse