Test immediately upon startup, no need to wait! The ME2000B explosion-proof clamp type grounding resistance tester adopts digital integration technology, which is lightweight and portable. It can measure in a wide area of 0.01 Ω -1200 Ω, with sound and light alarms indicating abnormalities. It can store 99 sets of data and safely and quickly solve grounding faults that cannot be measured by traditional methods. It is suitable for multiple fields such as power and telecommunications.

Product Details

ME2000B Explosion proof Clamp type Grounding Resistance TesterBriefly describe

Its performance is mainly reflected in:

There is no need to wait for a long self check after booting up, and you can directly enter the testing state after booting up.

Abandoning the traditional relay self checking method, adopting * algorithm and digital integrated processing technology.

Optimize the body design to solve the problem of bulky old products, which is more in line with the lightweight characteristics of handheld instruments.

Brand new exterior design, 6-key operation on the panel, with direct and convenient functions.

Added sound and light alarm function, with a "beep beep beep beep" sound indicating abnormality.

Add the function of identifying interference signals and providing prompts through a "beep beep beep beep" sound.

The measurement range has been expanded to 0.01 Ω -1200 Ω, making it more suitable for a wider range of scenarios.

It can store 99 sets of data, with lower power consumption and a maximum startup and operating current of less than 50mA.

ME2000B explosion-proof clamp type grounding resistance meter is suitable for measuring grounding resistance and loop resistance in fields such as power, telecommunications, meteorology, oil fields, construction, and industrial electrical equipment. When measuring a grounding system with a circuit, there is no need to disconnect the grounding lead or use auxiliary grounding electrodes, which is safe and fast. This instrument can measure the comprehensive value of grounding body resistance and grounding lead resistance, which can solve grounding faults that cannot be measured by traditional methods and is suitable for scenarios where traditional methods are limited.

ME2000B Explosion proof Clamp type Grounding Resistance TesterRange and accuracy

ME2000B 防爆型钳形接地电阻测试仪


ME2000B Explosion proof Clamp type Grounding Resistance TesterTechnical Specifications

resistance range 0.01Ω-1200Ω

Resistance resolution 0.001 Ω

99 sets of data storage

Sound and light alarm "beep beep beep beep" sound, press the AL key to turn on or off

The setting range of the alarm critical value is 1-199 Ω;

Power supply DC6V, No. 5 alkaline dry battery (LR6 1.5V × 4)

Working temperature and humidity -20 ℃ -55 ℃; 20%RH-90%RH

LCD display with 4-digit digital display, measuring 47mm in length and 28.5mm in width

Jaw size 65mm × 32mm

Jaw opening size 32mm

The weight of the clamp gauge is about 1100g (including battery), and the total weight is about 2.3kg (including accessories)

Clamp size: long clamp: length, width, and thickness 285mm × 85mm × 56mm

Protection level: Double insulation

Structural characteristics: Clamp CT

Automatic gear shifting

Single measurement time 0.5 seconds

Resistance measurement frequency >1KHz