Luheng Bio portable dissolved oxygen analyzer

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The LH-D701 portable dissolved oxygen analyzer from Luheng Biology is an intelligent analytical instrument that can be used for monitoring tap water sources, aquaculture, sewage treatment plants, and scientific research institutions to measure dissolved oxygen in water bodies. The instrument has a novel appearance and is easy to carry, suitable for on-site and field operations.

Product Details

Luheng Bio portable dissolved oxygen analyzerLH-D701

2.jpg


Performance characteristics:

1.Wide screen LCD display, operation prompts, easy to use
2.
Strong battery life, powered by lithium batteries
3.
Portable detection box
4.
Protection levelIP65

Technical parameters:

measurement range
Dissolved oxygen concentration:(0.00~ 20.00)mg/L
Dissolved oxygen saturation:(0.0~200.0%
temperature(0.0~60.0)℃(32~140) °F

resolution
Dissolved oxygen concentration:0.01mg/L
Dissolved oxygen saturation:0.1%
temperature0.1°C(0.1°F)

precision
Dissolved oxygen concentration:≤±0.30mg/L
Dissolved oxygen saturation:≤±3.0%
temperature≤0.3℃

Atmospheric pressure compensation: manual compensation(60.0~110.0)KPa

Salinity compensation: manual compensation(0.0~50.0) g/L

Temperature compensation: automatic compensation(0.0~40.0) ℃(32~ 104) °F

Standard electrode:SD02Dissolved oxygen electrode, cable3rice

Response time:≤45S

Power supply: lithium battery

Environmental protection level:IP65

Overall dimensions:165x75x51mm

Operation guide:

Installation of dissolved oxygen electrode:

1)Unscrew the electrode membrane head, clean the metal electrode and membrane head with pure water, and spin dry.

2)If three-quarters of the volume of electrolyte is injected into the membrane head, and then the membrane head is rotated onto the electrode, it is normal for some electrolyte to overflow during the tightening process.

3)Gently wipe off the spilled electrolyte and check if there are any bubbles in the electrolyte inside the film. If there are bubbles, reinstall the film head.

4)Gently press the membrane at the top of the membrane head with your hand to check if there is a gap between the membrane and the metal top. If there is, the membrane head needs to be reinstalled.

Electrode polarization:

1)Before the first use, polarize the electrode.

2)Open the instrument, connect the electrodes, and wait15Just a minute is enough.

Luheng Bio portable dissolved oxygen analyzerLH-D701Measurement of:

After cleaning the dissolved oxygen electrode with pure water, it is generally immersed in the measured solution. The instrument can be turned on for measurement, and the dissolved oxygen concentration is calculated while the instrument is in the measuring state“mg/L”Dissolved oxygen saturation“%”And electrode current value“nA”,Short buttonchoice/bit selectionSwitch the measurement status display.