HD-2001Low background multi-channel gamma spectrometer
Product Introduction
HD-2001Low background multi-channel gamma spectrometerMainly used for gamma spectroscopy measurement and analysis of building materials, soil, biological, geological samples, etc; Measurement and analysis of radon concentration in the air and soil; Rapid non-destructive testing of building materials; Analysis of radium (uranium), thorium, and potassium content in uranium geological samples; Artificial nuclide analysis software such as uranium, cesium, cobalt, iodine, etc. can be equipped according to user requirements
execution standard
GB/T 11743-2013 "Method for gamma spectroscopy analysis of radioactive nuclides in soil"
GB 6566-2010 "Radionuclides in Building Materials"
GB 50325-2010 "Code for Indoor Environmental Pollution Control of Civil Building Engineering" (2013 edition)
Features:
The actual application of HD2001 is a multi-channel y-spectrometer with 2048 channels
2. It has four major functions: powder sample measurement, non-destructive sample measurement, air hydrogen measurement, and soil hydrogen measurement
3. Full wave acquisition, no pulse leakage meter, with intelligent functions such as baseline correction and interference recognition. PMT output signal is straight through, waveform distortion free, and consistent resolution can be measured by placing C5 peak at any channel address within the measurement range
4. Maximum data pass rate:>333Kcps, high-speed transmission, no data leakage count
5. Visual measurement display, with functions such as signal waveform, baseline online monitoring, and real-time analysis of measurement results
6. Equipped with peak position monitoring function to ensure data collection quality
7. Comprehensive analysis and spectral analysis of characteristic spectral segments, full spectrum testing, multi spectral line comparison, and other functions can be used for scientific research, teaching, and technical training
8. Electric structure lead shielding room, saving time and effort when replacing samples

3、 Technical standards:

4、 Specific parameters“
Technical indicators:
1. Detector: Nal (Tl) crystal with a diameter of 75x75 mm
2. Total number of channels: 512, 1024, 2048, 4096 channels to choose from
Actual application channels: 2048 channels
3. Energy resolution:<7.5% (137C5)
4. Instrument background: ≤ 5.0cps (50keV~3MeV)
5. Differential nonlinearity:<0.05%
6. Integral nonlinearity:<0.10%
7. Stability: Relative error ≤ 1.0% (24h)
8. Pulse resolution: 500ns
9. Instrument detection limit
226Ra: 8.0 Bq/kg
232Th: 6.0 Bq/kg
40K:20.0 Bq/kg
222Rn (air): 6.0 Bq/m
222Rn (soil): 300.0 Bq/m
10. Highest data pass rate:>333 kcp511. Measurement uncertainty (sample radioactivity>37Bq/kg):<10% 12. Operating environment
Temperature: (+5~+40)
Humidity: ≤ 90%
13. Power supply: 220V (10%) 50HZ
14. Size and weight of lead shielding room:
(Ф580x850)mm; Wall thickness 100mm; weight approximately 900kg