Instrument configuration
Standard P100 high-pressure constant current pump UV100 UV detector 7725i manual injection valve Exformma Pronaos series chromatographic columns W100 workstation software. Digital computer intelligent fully controlled liquid chromatography system. Reciprocating double plunger parallel pump. More stable and reliable, with a longer service life. The appearance is novel, the operation is convenient, and it reflects humanization. Excellent performance indicators, comparable to popular foreign models. The optical system adopts precision positioning structure and thermal isolation installation technology, which has the characteristics of high accuracy, small offset, and short stability time. All components are manufactured using world-class CNC machining centers.
Technical indicators:
Basic technical specifications of high-pressure constant flow pump: Flow range: 0.001-9.999 mL/min Flow stability error: Sr≤0.3% RSD<0.06% Error in flow setting value: Ss≤2% Pressure linearity and accuracy: Display pressure error ≤± 0.5MPa (0-42MPa) Pressure pulsation: ≤ 0.1 Mpa Sealing pressure of pump: 42MPa Maximum working pressure: 42MPa (0.001-9.999mL/min) Dimensions: 450mm x 300mm x 160mm (length x width x height)
Basic technical specifications of UV detector: Wavelength range: 190-680nm Spectral bandwidth: 8nm Wavelength indication error: ≤ ±1nm Wavelength repeatability better than 0.1nm Baseline noise: ≤ 2 × 10-5AU (dynamic) Baseline drift: ≤ 2 × 10-4AU (dynamic) Minimum detection limit: 1 × 10-8g/mL (methanol solution of naphthalene) Dimensions: 450mm x 300mm x 160mm (length x width x height)
|